 Figure 3: Graph of pH versus volume

Tips

Except for opening and closing the stopcocks, do not touch 
the syringe during the titration; otherwise, the drop size may 
change significantly.

The drop rate must remain approximately constant (about 2 
drops per second) in order for the drop size to remain con-
stant.

Drop size depends on the exact position of the bottom stop-
cock. Therefore, the average drop size will be different for 
every titration. For each titration, re-enter the volume used 
and drop count values in the calculation.

Other Titrations

Titrations with different solutions can be performed using the 
same equipment and procedure. Other probes (such as conductiv-
ity or ORP) can be used in place of the pH probe.

Theory of Operation

The Drop Counter uses an 
infrared light source and a 
photodetector. Lenses 
spread the light into a 
“sheet” and refocus it 
onto the photodetector. 
When a drop of fluid 
passes through the Drop 
Counter, it partially 
blocks this sheet of light, and the photodetector registers a 
momentary decrease in light intensity.

The Drop Counter uses infrared light and ignores visible light. 
On power up, it automatically adjusts the light level for the best 
sensitivity. Use the Drop Counter away from direct sunlight or 
other sources of infrared that may interfere with it.

When the Drop Counter detects a falling drop, the green indicator 
light emitting diode (LED) flashes. If something stationary is 
blocking the photodetector, the indicator LED remains on.

Troubleshooting

Specifications

Technical Support

For assistance with any PASCO product, contact PASCO at:

For more information about the Drop Counter and the latest revi-
sion of this Instruction Sheet, visit the PASCO web site and enter 
PS-2117 in the Search window.

Limited Warranty

For a description of the product warranty, see the 

PASCO catalog.

Copyright

The PASCO scientific 012-08470D 

Drop 

Counter Instruction Sheet

 is copyrighted with all rights reserved. Permis-

sion is granted to non-profit educational institutions for reproduction of any 
part of this manual, providing the reproductions are used only in their labo-
ratories and classrooms, and are not sold for profit. Reproduction under 

Select volume calculation 

for horizontal axis

detector

collimating

lens

focusing 
lens

sheet of light

drop

infrared light 

source

 Figure 4: Diagram of Drop Counter

Problem

Cause

Solution

Indicator light flashes 
twice for a single drop. 
Drop Counter registers 
more drops than 
actually dispensed.

Fluid is splashing 
back into the drop 
window.

Adjust the position 
of the beaker, Drop 
Counter, or syringe 
to eliminate splash 
back

Indicator LED remains 
on constantly. Drop 
Counter does not 
register drops.

Lenses in drop 
window are dirty or 
wet.

Clean lenses with 
water and dry with a 
cotton swab or 
tissue.

Maximum drop rate

40 drops per second

Default sample rate

5 samples per second

Maximum sample rate

10 samples per second

Minimum object detection size

0.64 mm diameter

Enclosure

Polypropylene, water resistant, 
chemically resistant to mild acids 
and bases

Detector type

Infrared photodetector with 