Pasco Scientific OS-8501 Скачать руководство пользователя страница 9

scientific

5

012-02675B

Interferometer

Figure 5 ADJUSTING M

1

LASER

REFLECTED 
LASER BEAM

OPTICS BENCH

ALIGNMENT RAIL

VIEWING SCREEN

LASER BEAM

10 - 20 cm

Aligning the Interferometer

 NOTE:  This alignment procedure is for

those using a PASCO scientific Optics Bench.  If
you are not using an Optics Bench, tape a
straightedge to a flat level surface. The straight-
edge will provide a substitute for the alignment
rail of the optics bench.

Place the laser and the interferometer on the Optics
Bench, approximately 10 - 20 cm apart (Figure 5).
Be sure that the edges of both units are flush
against the alignment rail of the bench.  Place a
viewing screen as shown.  (A blank sheet of white
paper taped to the cover of a book provides a con-
venient screen.) Turn on the laser.

Loosen the thumbscrew that holds the beam-splitter
and rotate the beam-splitter so it is out of the beam
path of the laser as shown in Figure 5.  Then loosen
the thumbscrew that holds M

2

, the movable mirror.

Adjust the rotation of M

2

 so the laser beam is re-

flected directly back toward the aperture of the la-
ser.  (The reflected beam need not be at the same
height as the incident beam, but it should strike the
front panel of the laser along a vertical line through

the aperture.)  Hold M

2

 in position and tighten the

thumbscrew.

Rotate the beam-splitter so its surface is at an angle
approximately 45

o

 with the incident beam from the

laser (see Figure 6).  You will see two sets of laser
spots on the viewing screen, corresponding to the
two paths that the beam takes in reaching the
screen.  (Each path results in more than one laser
spot because of multiple reflections within the
beam-splitter.)  Adjust the beam-splitter so the two
sets of laser spots are as close as possible, then
tighten the thumbscrew to secure the beam-splitter.

Using the alignment screws, adjust the angle of M

1

until the two sets of laser spots are superimposed
on the viewing screen (the two brightest spots must
be superimposed).

Place the lens holder on the optical bench as shown
in Figure 7.  Be sure its edge is flush against the
alignment rail. Then place the 18 mm focal length
lens on the lens holder (it attaches magnetically).
Adjust the position of the lens on the holder so the
light from the laser, now spread out by the lens,
strikes the center of the beam-splitter.  If you have
performed the alignment correctly, you will see an
interference pattern of concentric rings on the view-
ing screen.  If the alignment is not just right, the
center of the fringe pattern may not be visible on
the screen.  Adjust the alignment screws on M

1

very slowly as needed to center the pattern.

Содержание OS-8501

Страница 1: ...Copyright February 1986 10 00 Instruction Manual and Experiment Guide for the PASCO scientific Model OS 8501 012 02675 10 91 Revision B Interferometer scientific MODEL OS 8501 INTERFEROMETER...

Страница 2: ......

Страница 3: ...t Needed Additional Equipment Recommended Theory of Operation 2 Interference Theory The Michelson Interferometer Operation The Interferometer 4 The Movable Mirror 4 Aligning the Interferometer 5 Exp 1...

Страница 4: ...amage caused by improper packing of the equipment for return shipment will not be covered by the warranty Shipping costs for returning the equipment after repair will be paid by PASCO scientific Copyr...

Страница 5: ...laser We recommend the PASCO 0 5 mW He Ne Laser Model OS 9171 but any low power laser that operates in the visible range will work For optimum ease of alignment the level of the beam should be 1 5 in...

Страница 6: ...an interference pattern He allowed a single narrow beam of light to fall on two narrow closely spaced slits Opposite the slits he placed a viewing screen Where the light from the two slits struck the...

Страница 7: ...or M2 the path length of one of the beams can be varied Since the beam traverses the path between M2 and the beam splitter twice moving M2 1 4 wavelength nearer the beam splitter will reduce the optic...

Страница 8: ...ontrolled and measured using the micrometer knob Each division of the knob corresponds to 1 micrometer 10 6 meter of mirror movement The Movable Mirror To measure the wavelength of light the movement...

Страница 9: ...rtical line through the aperture Hold M2 in position and tighten the thumbscrew Rotate the beam splitter so its surface is at an angle approximately 45o with the incident beam from the laser see Figur...

Страница 10: ...M1 ALIGNMENT SCREWS LENS 18 mm FOCAL LENGTH LENS HOLDER LASER VIEWING SCREEN 45 M1 ALIGNMENT SCREWS Figure 6 ALIGNING THE LASER SPOTS Figure 7 POSITIONING THE LENS Adjust the Beam Splitter and M1 to s...

Страница 11: ...nt in any mechanical system involving reversals in direction of movement By beginning with a full counterclockwise turn and then turning only counterclockwise when counting fringes you can eliminate b...

Страница 12: ...Then plug the banana plug of the vacuum chamber into the hole in the interferometer base between the fixed mirror and the beam splitter Adjust the alignment screws of the fixed mirror so the center of...

Страница 13: ...auge Of course your room pressure may vary slightly from 76 cm Hg If you have a more accurate way of measuring room pressure you can use that value to replace 76 cm Hg in the above equation Mark a ref...

Страница 14: ...difference between these values mi mf is just m the number of fringes you counted as you evacuated the chamber Therefore m 2 d i 2d f However i o ni and f o nf where ni and nf are the initial and fin...

Страница 15: ...rs no other maintenance is required If the mylar strip gets damaged or a severe shock causes the interferometer to lose its alignment return the unit to PASCO scientific for repair Replacement Parts T...

Страница 16: ...e the interferometer with a spectral or other monochromatic light source Tape two thin pieces of wire or thread to the sur face of the diffuser to form cross hairs Set up the equipment as shown below...

Страница 17: ...ting Technical Support Before you call the PASCO Technical Support staff it would be helpful to prepare the following infor mation If your problem is with the PASCO apparatus note Title and model numb...