Pasco Scientific OS-8501 Скачать руководство пользователя страница 11

scientific

7

012-02675B

Interferometer

Exp 1:  Measuring the Wavelength of Light

Theory

In many scientific and industrial uses of interferometers, a light source of a known wave-
length is used to measure incredibly small displacements - about 10

-6

 meters.  However, if

you know the distance of mirror movement, you can use the interferometer to measure the
wavelength of a light source.  In this experiment you will use the interferometer to measure
the wavelength of your laser light source.

Procedure

Align the laser and interferometer as described in the preceding section, so an interference
pattern of circular fringes is clearly visible on your viewing screen.

Adjust the micrometer knob so the lever arm is approximately parallel with the edge of the
interferometer base.  In this position the relationship between knob rotation and mirror
movement is most nearly linear.

Turn the micrometer knob one full turn counterclockwise.  Continue turning counterclock-
wise until the zero on the knob is aligned with the index mark.

NOTE:  Whenever you reverse the direction in which you turn the micrometer knob,
there is a small amount of give before the mirror begins to move.  This is called
mechanical backlash, and is present in any mechanical system involving reversals in
direction of movement.  By beginning with a full counterclockwise turn, and then turning
only counterclockwise when counting fringes, you can eliminate backlash in your
measurement.

If you are using a blank piece of paper as your viewing screen, make a reference mark on
the paper between two of the fringes.  You will find it easier to count the fringes if the
reference mark is one or two fringes out from the center of the pattern.

Rotate the micrometer knob slowly counterclockwise.  Count the fringes as they pass your
reference mark.  Continue until a predetermined number of fringes has passed your refer-
ence mark (count at least 20 fringes).  As you finish your count, the fringes should be in the
same position with respect to your reference mark as they were when you started to count.

Record d

m

, the distance that the movable mirror moved toward the beam-splitter as you

turned the micrometer knob.  Remember, each division on the micrometer knob corre-
sponds to one micron (10

-6

 meters) of mirror movement.

d

m

= _______________

Record m, the number of fringes that crossed your reference mark.

m = _______________

Calculate the wavelength of the laser light (

λ

 = 2d

m

/m).

Calculate the percentage difference between your measured value for the wavelength of the
laser light and the value recorded in the laser specifications.  (Check with your teacher for
the laser specifications.)

Содержание OS-8501

Страница 1: ...Copyright February 1986 10 00 Instruction Manual and Experiment Guide for the PASCO scientific Model OS 8501 012 02675 10 91 Revision B Interferometer scientific MODEL OS 8501 INTERFEROMETER...

Страница 2: ......

Страница 3: ...t Needed Additional Equipment Recommended Theory of Operation 2 Interference Theory The Michelson Interferometer Operation The Interferometer 4 The Movable Mirror 4 Aligning the Interferometer 5 Exp 1...

Страница 4: ...amage caused by improper packing of the equipment for return shipment will not be covered by the warranty Shipping costs for returning the equipment after repair will be paid by PASCO scientific Copyr...

Страница 5: ...laser We recommend the PASCO 0 5 mW He Ne Laser Model OS 9171 but any low power laser that operates in the visible range will work For optimum ease of alignment the level of the beam should be 1 5 in...

Страница 6: ...an interference pattern He allowed a single narrow beam of light to fall on two narrow closely spaced slits Opposite the slits he placed a viewing screen Where the light from the two slits struck the...

Страница 7: ...or M2 the path length of one of the beams can be varied Since the beam traverses the path between M2 and the beam splitter twice moving M2 1 4 wavelength nearer the beam splitter will reduce the optic...

Страница 8: ...ontrolled and measured using the micrometer knob Each division of the knob corresponds to 1 micrometer 10 6 meter of mirror movement The Movable Mirror To measure the wavelength of light the movement...

Страница 9: ...rtical line through the aperture Hold M2 in position and tighten the thumbscrew Rotate the beam splitter so its surface is at an angle approximately 45o with the incident beam from the laser see Figur...

Страница 10: ...M1 ALIGNMENT SCREWS LENS 18 mm FOCAL LENGTH LENS HOLDER LASER VIEWING SCREEN 45 M1 ALIGNMENT SCREWS Figure 6 ALIGNING THE LASER SPOTS Figure 7 POSITIONING THE LENS Adjust the Beam Splitter and M1 to s...

Страница 11: ...nt in any mechanical system involving reversals in direction of movement By beginning with a full counterclockwise turn and then turning only counterclockwise when counting fringes you can eliminate b...

Страница 12: ...Then plug the banana plug of the vacuum chamber into the hole in the interferometer base between the fixed mirror and the beam splitter Adjust the alignment screws of the fixed mirror so the center of...

Страница 13: ...auge Of course your room pressure may vary slightly from 76 cm Hg If you have a more accurate way of measuring room pressure you can use that value to replace 76 cm Hg in the above equation Mark a ref...

Страница 14: ...difference between these values mi mf is just m the number of fringes you counted as you evacuated the chamber Therefore m 2 d i 2d f However i o ni and f o nf where ni and nf are the initial and fin...

Страница 15: ...rs no other maintenance is required If the mylar strip gets damaged or a severe shock causes the interferometer to lose its alignment return the unit to PASCO scientific for repair Replacement Parts T...

Страница 16: ...e the interferometer with a spectral or other monochromatic light source Tape two thin pieces of wire or thread to the sur face of the diffuser to form cross hairs Set up the equipment as shown below...

Страница 17: ...ting Technical Support Before you call the PASCO Technical Support staff it would be helpful to prepare the following infor mation If your problem is with the PASCO apparatus note Title and model numb...

Страница 18: ......

Отзывы: