8. 

Calibration Period

:    

a.  Select the number of days in the first column and the number of tests in the 

second column that may be allowed to occur before calibration is needed. 

b. 

Select “LOCK” to lock the device if calibration is needed. 

c.  Default is 360 days, 700 tests and unlocked. 

 
 
 
 

Calibration 

 
PAS recommends all MARK V units be calibrated annually or whenever an accuracy 
test shows that BAC readings are no longer within tolerance limits.  Because precise 
calibration  is  crucial,  it  can  be  performed  only  by  PAS  Systems  International  or  by 
persons who have been properly trained for alcohol detector calibration.  (Calibration 
of  the  MARK  V 

must

  be  performed  using  a  NHTSA-approved  wet  bath  simulator  or 

certified dry gas standard.)  Successful completion of calibration should be recorded 
in a calibration log.   
 
NOTE: When using dry gas (PAS recommends 0.080% dry gas concentration) for 
calibration, the alcohol concentration printed on the label of the dry gas cylinder must 
be corrected to account for altitude when conducting tests at or above an altitude of 
250 feet above sea-level. Please refer to the Operating Manual for specific 
instructions for altitude corrections. 
 
 
 
 

 
CALIBRATION: (Using an approved Wet Bath Simulator or Certified Dry Gas 
with proper Regulator) 

 

1.  Select the Settings Icon 

 

and select “Cal.”.  Calibration can also be 

accessed under Advanced Settings. 

2.  Enter the password 119119. 
3.  Enter the concentration of the calibration solution/cylinder.  Accuracy should be 

to one decimal place.  The preset unit is mg/100mL.  For example, enter 80.0 
for 0.080% or enter 100.0 for a 0.100% solution.   
 
 
 
 
 

Содержание MARK V Alcovisor

Страница 1: ...r Advanced Settings Manual PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Copyright 2010 by PAS Systems International Inc All R...

Страница 2: ......

Страница 3: ...bration Passwords Default Passwords Calibration 119119 Advanced Settings 112112 The default passwords for advanced settings and calibration can be changed To change the passwords please refer to the P...

Страница 4: ...ly if the optional printer is available and input the number of copies to be printed Select Cal to have the Last Calibration Date print 5 Delete This option allows for the deletion of ALL saved test r...

Страница 5: ...or or certified dry gas standard Successful completion of calibration should be recorded in a calibration log NOTE When using dry gas PAS recommends 0 080 dry gas concentration for calibration the alc...

Страница 6: ...the MARK V to an approved wet bath simulator cylinder Use calibration adaptors between the mouthpiece and simulator or cylinder for an airtight fit if necessary Highlight Cal on the screen using the...

Страница 7: ...nician Only original MARK V parts may be used To keep the instrument clean periodically use a mild disinfectant and a soft cloth on the outside of the case DO NOT use alcohol to clean the unit Warrant...

Страница 8: ...MARK V Alcovisor PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 http www pasintl com Tel 540 372 3431 800 660 SNIF Printed 09 13...

Отзывы: