Advanced Settings 

 
To access the Advanced Settings menu: 

1.  Choose the Settings Icon 

 

2. 

Select “Advanced” 

3.  Enter the password 112112 

 
  

Advanced Settings: 

 

1. 

Blow

:  This icon under advanced settings allows the user to change the blow 

time (in seconds) and blow pressure (1 to 6, with 6 being the hardest).  These 
are set to optimal levels already and it is not recommended to change them. 

 

2. 

Default

:  Select 

“yes” under this icon to return instrument to Default settings.  

Note: Settings have been optimized; it is not recommended to return to Default 
settings.  
 

3. 

Unit

:  Set the measurement unit and BAC/BrAC ratio.  If the unit is changed, the 

instrument does not need to be recalibrated. 
 

4. 

Print

:  Select Auto Print for a result to be printed automatically (if the optional 

printer is available) and input the number of copies to be printed.  

Select “Cal.” 

to have the Last Calibration Date print. 
 

5. 

Delete

:  This option allows for the deletion of ALL saved test records stored in 

the MARK V. 

Once you select “Yes”, the test records cannot be restored. 

 

6. 

Calibration

:  The password to access the calibration option is 119119.  Please 

refer to the calibration instructions included in this manual for proper calibration 
procedures.  
 

7. 

Password

:  The default passwords for advanced settings and calibration can be 

changed.  Select “Cal” or “Advanced Setting” to change the password.  Enter 

the new password and select 

  

to confirm.  Screen will display “Password 

Setting Succeed”.   
 

NOTE:  Instrument requires a 6-digit password.  If less than six  
digits are used, password will not store properly. 
 
Warning:  If you change the password, write it down.  The new password 
must be sent in with your unit if it should need service.  

 
 

Содержание MARK V Alcovisor

Страница 1: ...r Advanced Settings Manual PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Copyright 2010 by PAS Systems International Inc All R...

Страница 2: ......

Страница 3: ...bration Passwords Default Passwords Calibration 119119 Advanced Settings 112112 The default passwords for advanced settings and calibration can be changed To change the passwords please refer to the P...

Страница 4: ...ly if the optional printer is available and input the number of copies to be printed Select Cal to have the Last Calibration Date print 5 Delete This option allows for the deletion of ALL saved test r...

Страница 5: ...or or certified dry gas standard Successful completion of calibration should be recorded in a calibration log NOTE When using dry gas PAS recommends 0 080 dry gas concentration for calibration the alc...

Страница 6: ...the MARK V to an approved wet bath simulator cylinder Use calibration adaptors between the mouthpiece and simulator or cylinder for an airtight fit if necessary Highlight Cal on the screen using the...

Страница 7: ...nician Only original MARK V parts may be used To keep the instrument clean periodically use a mild disinfectant and a soft cloth on the outside of the case DO NOT use alcohol to clean the unit Warrant...

Страница 8: ...MARK V Alcovisor PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 http www pasintl com Tel 540 372 3431 800 660 SNIF Printed 09 13...

Отзывы: