Advanced Settings
To access the Advanced Settings menu:
1. Choose the Settings Icon
2.
Select “Advanced”
3. Enter the password 112112
Advanced Settings:
1.
Blow
: This icon under advanced settings allows the user to change the blow
time (in seconds) and blow pressure (1 to 6, with 6 being the hardest). These
are set to optimal levels already and it is not recommended to change them.
2.
Default
: Select
“yes” under this icon to return instrument to Default settings.
Note: Settings have been optimized; it is not recommended to return to Default
settings.
3.
Unit
: Set the measurement unit and BAC/BrAC ratio. If the unit is changed, the
instrument does not need to be recalibrated.
4.
: Select Auto Print for a result to be printed automatically (if the optional
printer is available) and input the number of copies to be printed.
Select “Cal.”
to have the Last Calibration Date print.
5.
Delete
: This option allows for the deletion of ALL saved test records stored in
the MARK V.
Once you select “Yes”, the test records cannot be restored.
6.
Calibration
: The password to access the calibration option is 119119. Please
refer to the calibration instructions included in this manual for proper calibration
procedures.
7.
Password
: The default passwords for advanced settings and calibration can be
changed. Select “Cal” or “Advanced Setting” to change the password. Enter
the new password and select
to confirm. Screen will display “Password
Setting Succeed”.
NOTE: Instrument requires a 6-digit password. If less than six
digits are used, password will not store properly.
Warning: If you change the password, write it down. The new password
must be sent in with your unit if it should need service.