12 GB/IE/NI
Selecting the temperature unit
After switching on, the last temperature unit used will initially be selected.
Briefly press the °C / °F / SET button
11
to change the temperature units
7
(°C or °F).
The official temperature unit in Europe in °C.
Measuring temperature
Allow about 30 minutes for the product to acclimate to the conditions of the envi-
ronment.
Hold the Measure button
3
whilst aiming the laser pointer with 8 dot circle
1
at the surface you wish to measure.
The temperature is measured whilst pressing the Measure button; the display
5
will read „SCAN“.
After releasing the Measure button the last temperature measured will appear
in the display and the laser pointer with 8 dot circle will be deactivated.
The temperature determined is the average temperature of the area measured.
The area measured is circular. The diameter of the area measured „S“ is one
twelfth of the distance between the surface and the infrared sensor
2
„D“.
Below you will find some values as a guide:
S (for surface diameter):
10 mm
20 mm
30 mm
D (for distance):
120 mm
240 mm
360 mm