Main Menu 

SUMMARY: This menu allows the setting of various parameters as they relate to the process being monitored.  Use “Menu” 
to scroll through the menu, the up/down arrow keys to scroll through choices, and “Enter” to select the choice.   
 
NOTE: The analog output of the Temperature is over a range of 0‐100⁰C and is not adjustable. 

 

MAIN SCREEN: Displays Conductivity (CD:), Temperature in °C. in line 1, and an analog bar graph of % of full scale 
(conductivity) along with the actual % value.  

 

CONCENTRATION UNITS: Use the up / down arrows to select between µS (micro siemens), mS (milli siemens), ppm KCl 
(parts / million potassium chloride), ppm NaCl (parts / million sodium chloride), ppm 442 (parts / million 442 solution). 
Default = µS 

 

Range Hi: Use to set the upper range limit of the Analog output signal (100% full scale).  Allowable maximums are 
200000 µS, 200 mS, 133566 ppm KCl, 12622 NaCl, 203794 ppm 442.  Default = 200000 µS 

 

RANGE LO: Use to set lower range limit of the Analog output signal (0.0% of full scale).  Default minimum is 0.0 
regardless of scale. 

 

TEMP REF: Reference Temperature. It can be changed if needed; all sensors are pre‐calibrated using a reference 
temperature of 25.0 °C. Default = 25.0 

 

TEMP COMP: Temperature Compensation, Turn ON or OFF, Default = ON. 

 

COLD ROOM MODE: Improves performance in rooms at approximately 4 °C. (Anything considerably colder than 
ambient) Default – OFF. 

 

TEMP COEFF: Temperature Coefficient, used when turned on above, 0.0 – 5.00 % / °C.  Default is 1.90 % / °C. 

 

Serial: Displays the RS‐232 serial output parameters.  Only baud rate can be changed.  Available rates are 1200, 2400, 
4800, 9600, 19200, 28800, 38400 and 57600.  Default = 9600 

 

PRINT TIME: Sets the data output interval, 1 to 60 sec.  Default = 5 sec. 

 

HI – ALARM: Used to set High Conductivity Alarm, value should be within High and Low Range set above.  Default = 0 µS 

 

LO – ALARM: Used to set Low Conductivity Alarm, value should be within High and Low Range set above. Default = 0 µS 

 

HI – ALARM: Set High Temperature Alarm, Max 50 °C. Default = 0.0 °C. 

 

LO – ALARM: Set Low Temperature Alarm, Max 50 °C. Default = 0.0 °C. 

 

PASSWORD: A numerical Password may be entered, with 0 being the default value, and equivalent to No Password. 
Once entered, the Password is required to enter any menu on the SciCon II Monitor. Call Parker Technical Support if 
the password must be reset. 

 

FACTORY RESET: Used to restore the SciCon II Conductivity Monitor to the factory default values.  Change No to Yes 
using the arrow keys and press “Enter” to reset, change No to Yes on the “Are You Sure?” screen and press “Enter” 
again to finish the process. 

 

SET TIME/DATE…  Press Enter to modify the Time Date Stamp used by the SciCon II internal clock.  Press Menu to 
bypass and return to the top of the menu or Exit to leave the menu. 

 

TIME: Use the up / down arrows to change the time in the system. The format is 24 hour time, HH:MM. Press Enter to 
save. 

 

MONTH: Use the up / down arrows to change the Month. Press Enter to save. 

 

DAY OF MONTH: Use the up / down arrows to change the Day of the Month. Press Enter to save.

 

 

YEAR: Use the up / down arrows to change the Year.  Press Enter to save.

 

Содержание SciCon II

Страница 1: ...SciCon II Conductivity Temperature Monitor Single Use Sensor Installation Operating Maintenance Instructions ENGINEERING YOUR SUCCESS...

Страница 2: ...publication is for general information only and customers are requested to contact your local Bioscience Filtration Sales and or Technical Support contact for detailed information and advice on a pro...

Страница 3: ...iCon II Monitor Hardware Overview 13 Front Panel Data Entry Display 14 Back Panel Interface Options 15 SciCon II Single Use Conductivity Temperature Sensors 18 Part B SciCon II Monitor Software Overvi...

Страница 4: ...LLING SERVICING CLEANING OR REMOVING THE FUSE FAILURE TO DO SO COULD RESULT IN BODILY HARM AND OR PROPERTY DAMAGE CAUTION OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE DEVICES WARNING ONLY...

Страница 5: ...T DE RETIRER LE FUSIBLE NE PAS OBSERVER CES PR CAUTIONS RISQUERAIT DE CAUSER DES BLESSURES CORPORELLES OU ET D ENDOMMAGER L QUIPEMENT PRUDENCE SOYEZ PRUDENT LORSQUE VOUS MANIPULEZ DES APPAREILS SENSIB...

Страница 6: ...7 784 2234 DH3 2SF England phone 1 805 604 3400 phone 44 0 191 4105121 fax 1 805 604 3401 fax 44 0 191 4105312 email bioscience na parker com email bioscience emea parker com www parker com bioprocess...

Страница 7: ...Con II are observed See SciCon II Specifications The protection provided by this equipment may be impaired if the SciCon II is used in a manner or for purposes not specified by the manufacturer Parker...

Страница 8: ...ables CBL 007 1 power supply cord with 3 adapter clips US EU and UK versions Set up 1 Connect the power supply using the applicable adapter clip US is shown 2 Connect the sensor cables to monitor port...

Страница 9: ...mphenol connector I O Port 2 Output Female DB25 four TTL outputs for alarms four 4 20 ma Analog output signals 18 bit resolution I O Port 3 Printer PC DB9FM RS 232 output I O Port 3 Serial I O Female...

Страница 10: ...omponent free The pins have not been tested to USP 88 Class VI All other fluid path components are USP 88 Class VI Sensor Type 4 Electrode Conductivity Cell gold plated pins Conductivity Range 1 S cm...

Страница 11: ...DY button Connect a SciCon II Sensor using the cable provided Use enclosed USB cable to connect with a PC to automatically create a USB Serial Device with an assigned Com Port in the Device Manager Go...

Страница 12: ...as both an xls file and a txt file Files are saved at C Serial Data Loggers by Monitor type All files can be readily turned into PDFs NOTE The charts are do not adjust their range automatically You mu...

Страница 13: ...wall adapter as a power supply The single use SciCon II Sensors utilize four gold plated electrodes in a polysulfone sensor body available in five sizes Luer 3 8 or 1 2 hose barb 3 4 and 1 Ladish San...

Страница 14: ...keys whose function does not change These keys are used for basic control and programming of the SciCon II The basic key definitions are MENU Allows entry into the parameter settings menu ENTER Accept...

Страница 15: ...ms Hi Low Conductivity Hi Low Temperature TTL outputs provide a 5 VDC control signal carry minimal current See Pin out on next page Printer PC RS232 output Female DB9 A printer cable P N 080 096 is re...

Страница 16: ...is needed in the loop connect pin 3 Vpwr to pin 16 4 20 and connect the output signal cable to pin 4 4 20 and pin 17 Gnd Analog outputs 3 and 4 are not utilized in the SciCon II TTL Outputs Hi Conduct...

Страница 17: ...17 The use of an end user supplied analog loop power source is presented below The SciCon II only uses Analog 1 and 2 Please note the black lines on the drawing above are internal connections...

Страница 18: ...hetic origin and animal derived material free to the best of our knowledge All other fluid path components are animal derived component free The pins have not been tested to USP 88 Class VI All other...

Страница 19: ...and 442 are also available the high and low analog range settings as well as the temperature reference and coefficient The user may also adjust the baud rate and print time and configure user defined...

Страница 20: ...AL BR 9600 WL 8 SB 2 PT N PRINT TIME 5 SEC HI ALARM 0 S ON OFF 0 00 5 00 C 1200 4800 9600 19200 28800 38400 57600 Menu Menu Menu Menu Menu 1 60 Seconds SEE HI RANGE FOR LIMIT MINIMUM 0 Menu Menu Menu...

Страница 21: ...00 C Default is 1 90 C Serial Displays the RS 232 serial output parameters Only baud rate can be changed Available rates are 1200 2400 4800 9600 19200 28800 38400 and 57600 Default 9600 PRINT TIME Set...

Страница 22: ...ivity T Temperature C LC Low Conductivity CF Calibration Factor HT Hi Temperature TO Temperature Offset LT Low Temperature CCF Custom Calibration Factor CTO Custom Temperature Offset CD Conductivity V...

Страница 23: ...23 Appendix A Sensor Performance Data...

Страница 24: ...24 Notes...

Страница 25: ...Co Durham DH3 2SF England phone 44 0 191 4105121 fax 44 0 191 4105312 email bioscience emea parker com www parker com bioscience Parker Hannifin Corporation Bioscience Filtration 2340 Eastman Avenue...

Отзывы: