SciCon II
™
Conductivity/Temperature Monitor & Single-Use Sensor
Installation, Operating & Maintenance Instructions
ENGINEERING
YOUR
SUCCESS.
Страница 1: ...SciCon II Conductivity Temperature Monitor Single Use Sensor Installation Operating Maintenance Instructions ENGINEERING YOUR SUCCESS...
Страница 2: ...publication is for general information only and customers are requested to contact your local Bioscience Filtration Sales and or Technical Support contact for detailed information and advice on a pro...
Страница 3: ...iCon II Monitor Hardware Overview 13 Front Panel Data Entry Display 14 Back Panel Interface Options 15 SciCon II Single Use Conductivity Temperature Sensors 18 Part B SciCon II Monitor Software Overvi...
Страница 4: ...LLING SERVICING CLEANING OR REMOVING THE FUSE FAILURE TO DO SO COULD RESULT IN BODILY HARM AND OR PROPERTY DAMAGE CAUTION OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE DEVICES WARNING ONLY...
Страница 5: ...T DE RETIRER LE FUSIBLE NE PAS OBSERVER CES PR CAUTIONS RISQUERAIT DE CAUSER DES BLESSURES CORPORELLES OU ET D ENDOMMAGER L QUIPEMENT PRUDENCE SOYEZ PRUDENT LORSQUE VOUS MANIPULEZ DES APPAREILS SENSIB...
Страница 6: ...7 784 2234 DH3 2SF England phone 1 805 604 3400 phone 44 0 191 4105121 fax 1 805 604 3401 fax 44 0 191 4105312 email bioscience na parker com email bioscience emea parker com www parker com bioprocess...
Страница 7: ...Con II are observed See SciCon II Specifications The protection provided by this equipment may be impaired if the SciCon II is used in a manner or for purposes not specified by the manufacturer Parker...
Страница 8: ...ables CBL 007 1 power supply cord with 3 adapter clips US EU and UK versions Set up 1 Connect the power supply using the applicable adapter clip US is shown 2 Connect the sensor cables to monitor port...
Страница 9: ...mphenol connector I O Port 2 Output Female DB25 four TTL outputs for alarms four 4 20 ma Analog output signals 18 bit resolution I O Port 3 Printer PC DB9FM RS 232 output I O Port 3 Serial I O Female...
Страница 10: ...omponent free The pins have not been tested to USP 88 Class VI All other fluid path components are USP 88 Class VI Sensor Type 4 Electrode Conductivity Cell gold plated pins Conductivity Range 1 S cm...
Страница 11: ...DY button Connect a SciCon II Sensor using the cable provided Use enclosed USB cable to connect with a PC to automatically create a USB Serial Device with an assigned Com Port in the Device Manager Go...
Страница 12: ...as both an xls file and a txt file Files are saved at C Serial Data Loggers by Monitor type All files can be readily turned into PDFs NOTE The charts are do not adjust their range automatically You mu...
Страница 13: ...wall adapter as a power supply The single use SciCon II Sensors utilize four gold plated electrodes in a polysulfone sensor body available in five sizes Luer 3 8 or 1 2 hose barb 3 4 and 1 Ladish San...
Страница 14: ...keys whose function does not change These keys are used for basic control and programming of the SciCon II The basic key definitions are MENU Allows entry into the parameter settings menu ENTER Accept...
Страница 15: ...ms Hi Low Conductivity Hi Low Temperature TTL outputs provide a 5 VDC control signal carry minimal current See Pin out on next page Printer PC RS232 output Female DB9 A printer cable P N 080 096 is re...
Страница 16: ...is needed in the loop connect pin 3 Vpwr to pin 16 4 20 and connect the output signal cable to pin 4 4 20 and pin 17 Gnd Analog outputs 3 and 4 are not utilized in the SciCon II TTL Outputs Hi Conduct...
Страница 17: ...17 The use of an end user supplied analog loop power source is presented below The SciCon II only uses Analog 1 and 2 Please note the black lines on the drawing above are internal connections...
Страница 18: ...hetic origin and animal derived material free to the best of our knowledge All other fluid path components are animal derived component free The pins have not been tested to USP 88 Class VI All other...
Страница 19: ...and 442 are also available the high and low analog range settings as well as the temperature reference and coefficient The user may also adjust the baud rate and print time and configure user defined...
Страница 20: ...AL BR 9600 WL 8 SB 2 PT N PRINT TIME 5 SEC HI ALARM 0 S ON OFF 0 00 5 00 C 1200 4800 9600 19200 28800 38400 57600 Menu Menu Menu Menu Menu 1 60 Seconds SEE HI RANGE FOR LIMIT MINIMUM 0 Menu Menu Menu...
Страница 21: ...00 C Default is 1 90 C Serial Displays the RS 232 serial output parameters Only baud rate can be changed Available rates are 1200 2400 4800 9600 19200 28800 38400 and 57600 Default 9600 PRINT TIME Set...
Страница 22: ...ivity T Temperature C LC Low Conductivity CF Calibration Factor HT Hi Temperature TO Temperature Offset LT Low Temperature CCF Custom Calibration Factor CTO Custom Temperature Offset CD Conductivity V...
Страница 23: ...23 Appendix A Sensor Performance Data...
Страница 24: ...24 Notes...
Страница 25: ...Co Durham DH3 2SF England phone 44 0 191 4105121 fax 44 0 191 4105312 email bioscience emea parker com www parker com bioscience Parker Hannifin Corporation Bioscience Filtration 2340 Eastman Avenue...