Operation

Preparation for Operation

 

•  Check items before power-on.  When installation and wiring are completed, be sure to check the  

 

    following points before power on.

 

Wiring:

 

Step 1  

The power supply and the output terminals are properly connected.

 

  Step 2  

Check the grounding.

 

  Step 3  

Detector and converter are paired together properly.

 

Supply voltage matches the specification

  Installation of the detector:
  Step 1  

The inlet and outlet cables are properly connected to the in and out converter connectors

 

  Step 2  

The flow direction arrow represents the correct flow direction.

  •  Fill the detector with the fluid and shut off the supply valve. After all flow has stopped check for leaks.

    Remove bubbles if present.

 Power-on:
  Step 1 

Turn on the power supply to converter.

  Step 2

     After turning on the power supply, warm up the converter for about 15 minutes.

 Operation:
  Step 1  

Start fluid flow and begin measurement.

 

  Step 2  

If fluid flow is not travelling in the same direction as the flow arrow the flow-rate output signal will  

   

not 

increase.

 

  Step 3  

Confirm that when flow is increased the flow rate output signal increases.

 

Zero Adjustment

Run the fluid to purge the bubbles from the detector body. Close valves to stop flow through the detector.  
The detector must remain full of fluid.  Push the ZERO ADJ switch on the panel for 2 seconds or more to 
start ZERO point adjustment.  Zero point adjustment requires about 10 seconds. It reverts automatically to 
measuring mode after the adjustment. During ZERO adjustment, the ZERO/AGC LED lamp blinks at the 
interval of 1 second.  Output signal should read 0%. If flow does not read 0 +/- 1%, repeat if necessary. 
Setting of this flow rate value will be the cutoff for all valves under this level.

After the proper wiring and installation of this device is completed you will be able to apply power and acquire 
a flow rate signal.

Power-on

Содержание PFM1 Series

Страница 1: ...Division Chemical Delivery Products 7075 East Southpoint Road Tucson Arizona 85756 phone 520 574 2600 fax 520 574 2700 www parker com veriflo Mobile App m parker com veriflo Installation and Operation...

Страница 2: ...14 Troubleshooting 15 Safety Guide and Installation and Operating Instructions available at www parker com veriflo WARNING Indicates that improper use or operation may result in serious injury CAUTIO...

Страница 3: ...in 100 mm Installation Hole Rotatable 3 878 in 98 5 1 mm 5 354 in 136 1 mm 0 375 in 9 53 mm 0 249 in 6 33 mm C D d 0 2 in 4 m m 3 15 in 80 1 mm IN OUT 3 937 in 100 mm Installation Hole Rotatable 3 78...

Страница 4: ...C 20mA or less A OUT Amp Out 4 20mA DC Load resistance 500 or less Communication RS 485 Protocol MODBUS Up to 32 units connectable including host EMC Compliance EMI EN55011 2007 EMS EN61326 1 2006 RoH...

Страница 5: ...Alarm output 1 3 COM Common for AL1 and AL2 4 FG Grounding 5 0 V Voltage of power supply 24 V DC 6 24V No Terminal Spec Name Description 7 RS485 RS485 Communication 8 P OUT Pulse output 9 A OUT Curren...

Страница 6: ...tor must be mounted for ease of draining and cleaning Install the flow detector with inlet lower than the outlet The arrow on side of Flow detector shows flow direction Make sure the arrow matches the...

Страница 7: ...moval Step 1 Connect flow detector coaxial cables with BNC connectors to converter Step 2 Confirm that the converter body and the terminal block plug socket are firmly fixed with the lock arm of the m...

Страница 8: ...t ZERO point 2 ADRS address switch Set addresses and communication parameters of RS 485 For setting address 25 fit the arrow of x10 dial address switch to 2 and the arrow of x1 dial address switch to...

Страница 9: ...bout 15 minutes Operation Step 1 Start fluid flow and begin measurement Step 2 If fluid flow is not travelling in the same direction as the flow arrow the flow rate output signal will not increase Ste...

Страница 10: ...graduation to set Measuring value less than the set value is deemed as 0 automatically Burnout Burnout It determines analog output value when output is in fault Hi Hysteresis High hysteresis The gap...

Страница 11: ...F S 10 F S contact type NC P Out Frequency Fault Fault contact type NC MFG No F08 XXXXX XX TAG No FLUID XXX FULL SCALE FLOW RANGE 300mL min SER No DETECTOR 0406XXXXXX SER No CONVERTER 06XXXXXXXX DEVI...

Страница 12: ...2 Transmission mode RTU MODE Parity Even number Odd number No parity Stop bit 1 or 2 Mode and Byte Composition You can monitor flow rate with the PAFC 100 through the communication protocol You may a...

Страница 13: ...r settings not defined Address switch Contents 80 0 ms 81 5 ms 82 10 ms 83 20 ms 84 30 ms 85 40 ms 86 50 ms 73 None Stop bit 1 Other settings not defined CHANGING COMMUNICATION PARAMETERS BACK TO DEFA...

Страница 14: ...me value as the previous setting Ultrasonic flowmeter requires minimal daily maintenance Checking tube joints and connections Check for leakage around connection points Check for any loose nut Tighten...

Страница 15: ...g to the malfunction Possible causes are listed below for each trouble phenomenon P Out High Alarm Low Alarm Communication Output A Out Error LED Alarm LED ON 1 OFF ON 0 mL min Keep previous values 2...

Страница 16: ...rect All LEDs blinking vaguely Turn off the mains and turn it on again after 10 second break Operating in normal PCB might be damaged END NO YES NO NO NO YES YES YES Troubleshooting A If LED display i...

Страница 17: ...ector properly Does display appear within 30 seconds Remove bubbles from detector Does ERROR LED disappear within 30 seconds Air bubbles adhered to the partition of the upper part of the sensor Turn o...

Страница 18: ...e Contact Technical Support Specialist Perform loop check Is flow direction correct Correct flow direction Correct combination of detector and converter Is detector size set correctly at converter Mak...

Страница 19: ...TART Are BNC connectors connected securely Connect securely Do foreign materials adhere to detector Remove the foreign materials Check detector visually Is detector working Contact Technical Support S...

Страница 20: ...top to bottom flow and horizontal piping Pressure fluctuation or vibrations due to the pump Stop the flow completely by closing both upstream and downstream valves especially for the small size applic...

Страница 21: ...g Reduce pulsation flow or relocate meter to the place without pulsation flow Increase damping factor by disregarding response speed if possible Bubbles contained in the liquid Eliminate the bubbles b...

Страница 22: ...rminals reading less than 100 Is detector filled with liquid Arrange the pipe to be fully filled with the liquid Bottom to top flow direction is preferable Pay attention for top to bottom flow and hor...

Страница 23: ...Notes 23...

Страница 24: ...of Issue 06 2013 2009 Parker Hannifin Corporation Parker Hannifin Corporation Veriflo Division Chemical Delivery Products 7075 East Southpoint Road Tucson Arizona 85756 phone 520 574 2600 fax 520 574...

Отзывы: