STARTING THE GENERATOR

1  

Inspect all of the system connection points and verify that they are secure.

2  

With both the inlet and outlet ball valves of the buff er vessel closed, open the ball valve on the air inlet port to allow the compressed  

 

air into the generator.

3  

Switch the electrical power on to the generator and wait whilst it runs through the controller initialisation routine.

4  

If the generator was in standby mode when the electrical power was removed it will default 

 

to standby mode on completion of the initialisation routine.

5  

Press            to initiate the start up routine.

 

If the start clean up option is enabled the generator will run through the Rapid Cycle before 

 

opening the buff er valve and the N outlet valve. Th

  e clean up cycle, which takes approximately 

 

160 seconds, is designed to clean the CMS bed of impurities, bring the generator up to production purity more rapidly, and prevent  

 

poor quality gas fl owing into the buff er. If the generator was running when the electrical power was removed (e.g. power failure) it   

 

will automatically run through a start up cycle (if enabled) and then commence normal operation. Wait until this cycle is complete  

 

and the menu displays “Running”. Th

  is may take several minutes for ppm generators.

6  

Partially open the ball valve on the inlet to the buff er vessel to allow it to pressurise slowly. When the pressure gauge on the buff er   

 

vessel reads within 0.5 barg (7psig) of the inlet pressure, check for leaks in the buff er vessel inlet piping and then fully open the ball  

 valve.

7  

Open the ball valve on the outlet of the buff er vessel and check for leaks in the piping between the vessel and the generator.

8  

Open the ball valve on the Nitrogen outlet.

 

Note: If the purity of the gas is not within specifi cation it will be vented to atmosphere through a vent solenoid within the generator  

 

and not delivered to the application. When the required purity is achieved the gas will be delivered to the application.

22

N2 Compact Nitrogen Gas Generator - User Guide. ©2020.

      

      
       Standby

       Standby

 O2 = 5.00%  

 O2 = 5.00%  
 Rapid Cycle

 Rapid Cycle

STOPPING THE GENERATOR

1  

Close the ball valve on the N2 Outlet port.

2  

Press          to initiate the shutting down sequence.

 

Th

  e generator will complete the current cycle and then exhaust both beds. Th

  is may take several 

 

minutes particularly on ppm generators.

3  

When the generator is depressurised it will revert to standby mode.

      

      
       Standby

       Standby

      

      
    Shutting Down

    Shutting Down

 O2 = 5.00%  

 O2 = 5.00%  
 Shutting Down

 Shutting Down

Содержание N2 NITROSource Compact N2C-2

Страница 1: ...NITROGEN GAS GENERATOR NITROSource Compact USER GUIDE 17 651 1000 01 20 Rev Preventative Maintenance Guide 17 651 1100 User Guide EN N2C 2 N2C 4 N2C 6 N2C 8 N2...

Страница 2: ...e Generator Stopping the Generator Fault Log Customer Settings Overview of the Controls Menu Interface Hour Meters Generator Supply Dryer Supply Purge Economy Remote Switching Overview of the Equipmen...

Страница 3: ...let Parameters Environmental Parameters Port Connections Electrical Parameters Packed Weights and Dimensions Troubleshooting Schematics Generator Weights and Dimensions Declaration of Conformity Techn...

Страница 4: ...04 N2 Compact Nitrogen Gas Generator User Guide 2020 SAFETY...

Страница 5: ...ent is high such as a confined space or poorly ventilated room the use of oxygen monitoring equipment is advisable Nitrogen is not a poisonous gas but in a concentrated form there is a risk of asphyxi...

Страница 6: ...AMMABLE COMPRESSED GAS Wear ear protection Pressurised components on the system Remote control Generator may start automatically without warning Conformit Europ enne Dispose in accordance with local r...

Страница 7: ...07 N2 Compact Nitrogen Gas Generator User Guide 2020 RECEIVING AND INSPECTING THE EQUIPMENT...

Страница 8: ...ase inform the delivery company immediately and contact your local Parker office STORAGE The equipment should be stored within the packing crate in a clean dry environment If the crate is stored in an...

Страница 9: ...ng the generator to the base of the crate 02 Lift the generator from the pallet using suitable slings and an overhead crane 03 Carefully move the generator to its final location using a forklift truck...

Страница 10: ...10 N2 Compact Nitrogen Gas Generator User Guide 2020 OVERVIEW OF THE EQUIPMENT...

Страница 11: ...e N2 Outlet port to buffer G Column B pressure gauge Air inlet pressure gauge Mains supply cable gland N2 inlet port from buffer vessel G Air Inlet port G N2 Outlet pressure gauge Cable glands N2 Outl...

Страница 12: ...12 N2 Compact Nitrogen Gas Generator User Guide 2020 LOCATING THE EQUIPMENT...

Страница 13: ...metre of compressed air the particulate count should not exceed 400 000 particles in the 0 1 0 5 micron size range 6 000 particles in the 0 5 1 micron size range and 100 particles in the 1 5 micron s...

Страница 14: ...14 N2 Compact Nitrogen Gas Generator User Guide 2020 INSTALLATION AND COMMISSIONING...

Страница 15: ...Generator User Guide 2020 Warning Only competent personnel trained qualified and approved by Parker should perform installation commissioning service and repair procedures REF REF REF REF DESCRIPTION...

Страница 16: ...NB 16mm ID piping between the ball valve and the outlet port of the buffer vessel The AOP010 filter provided should be installed in this line Follow the installation instructions provided with the fi...

Страница 17: ...te switching is activated in the customer settings menu MODBUS For MODBUS communicatin setup details refer to publication 17 650 0012 6 12 mm 3 7 mm 3 7 mm 3 7 mm 3 7 mm L Fuse terminal for the phase...

Страница 18: ...itional information on installation requirements If the pre treatment dryer is fitted with a purge economy feature it may be controlled using the volt free relay contacts on JP17 The relay is energise...

Страница 19: ...D is set to 6 Note The oxygen purity setting of the generator is marked on the rating plate The table below shows the correlation between the purity settings of the generator and the output current Th...

Страница 20: ...20 N2 Compact Nitrogen Gas Generator User Guide 2020 OPERATING THE GENERATOR...

Страница 21: ...SCRIPTION DESCRIPTION DESCRIPTION DESCRIPTION Start key Operating status indicator Green Running Yellow Starting up Shutting down Red Standby Menu navigation keys Service Fault indicator Yellow Servic...

Страница 22: ...l this cycle is complete and the menu displays Running This may take several minutes for ppm generators 6 Partially open the ball valve on the inlet to the buffer vessel to allow it to pressurise slow...

Страница 23: ...ime in hours that the generator has been operating in Economy mode O2 5 00 O2 5 00 Shutting Down Shutting Down 1 Hour meters 1 Hour meters 2 Fault log 2 Fault log 2 1 Fault 10 2 1 Fault 10 2 1 Fault 1...

Страница 24: ...n the system Note Any faults that are active when the power is switched off and remain active when the power is re applied will cause a new entry to be added into the fault log 2 1 Fault 1 Active 2 1...

Страница 25: ...mpact Nitrogen Gas Generator User Guide 2020 3 1 O2 Alarm 3 1 O2 Alarm Over ride 0 Over ride 0 3 2 O2 Calibration 3 2 O2 Calibration 5 00 5 00 3 3 O2 Alarm Level 3 3 O2 Alarm Level 5 05 5 05 3 4 O2 Al...

Страница 26: ...se The generator will continue to cycle as normal without delivering gas to the application If the back pressure is maintained for an additional 5 minutes the generator will stop cycling and enter Eco...

Страница 27: ...27 N2 Compact Nitrogen Gas Generator User Guide 2020 SERVICING THE GENERATOR...

Страница 28: ...hs 24 Months Generator Check status indicators located on the control panel Generator Check regulated outlet pressure Generator Check O2 purity System Check filter drains System Check inlet air qualit...

Страница 29: ...P010AO element Catalogue No s Description Contents M24 PPM 0002 Kit 02 Cell PPM 02 Cell PPM M24 PCT 0002 Kit 02 Cell 02 Cell Catalogue No s Description Contents M24 N2C 0001 Kit 24 Month N2C valve ove...

Страница 30: ...retaining screw 06 of the assembly and remove Cut the tie wrap and unscrew the oxygen sensor Remove the used cell and replace with the new one Note Ensure to wear gloves when handling the new cell to...

Страница 31: ...tering the calibrated level USING A CALIBRATED GAS SUPPLY Select menu 3 1 and enable the O Alarm Over ride Connect the gas supply to the O Analyser calibration port 1 on the side of the generator Loca...

Страница 32: ...ul the original reading from the analyser will be loaded Should this occur repeat the above steps On completion of the calibration return the ball valves back to their original position and remove the...

Страница 33: ...Note A strap wrench may be required for the removal of the 050 055 filter bowl Close the ball valves located on the inlet and outlet ports of the filter and depressurise it by opening the manual drai...

Страница 34: ...cated in the filter head with the new O ring provided Refit the filter bowl and head ensuring that the threads are fully engaged 01 and the locking details are aligned 02 Note To ensure that the bowl...

Страница 35: ...35 N2 Compact Nitrogen Gas Generator User Guide 2020 TECHNICAL SPECIFICATION...

Страница 36: ...trol system automatically switches the bed to regenerative mode All of the contaminants are vented from the CMS and a small portion of the outlet nitrogen gas is expanded into the bed to accelerate th...

Страница 37: ...8 11 0 13 3 15 3 17 1 N2C 8 M3 HR 3 38 5 89 10 24 14 86 18 01 24 02 29 33 33 93 37 81 CFM 2 0 3 5 6 0 8 7 10 6 14 1 17 3 20 0 22 3 Air to N2 N2C 2 to N2C 8 7 0 5 3 3 9 3 2 3 0 2 5 2 3 2 1 2 0 Outlet P...

Страница 38: ...e generator alone and does not take in to account any pre treatment dryer connected to the dryer supply terminals of the generator The dryer is fed directly from the generator supply PACKED WEIGHTS AN...

Страница 39: ...4 19 207 5 8 17 237 5 9 35 98 216 1 N2C 4 1040 40 9 450 18 628 27 72 375 14 76 466 5 18 37 406 5 16 681 5 26 83 753 5 29 66 868 5 34 19 207 5 8 17 237 5 9 35 145 319 7 N2C 6 1040 40 9 450 18 796 31 34...

Страница 40: ...erform the required service High Oxygen concentration Defective Oxygen cell Replace the oxygen cell Leak in system piping Open the access door and check all connection points for leaks Repair as neces...

Страница 41: ...ALY Corsico MI Tel 39 02 45 19 21 parker italy parker com JP JAPAN Tokyo Tel 81 0 3 6408 3901 KR SOUTH KOREA Seoul Tel 82 2 559 0400 KZ KAZAKHSTAN Almaty Tel 7 7272 505 800 parker easteurope parker co...

Страница 42: ...rformance is maintained Air quality continues to meet international standards Protection of downstream equipment personnel and processes Low operational costs Increased productivity and profitability...