icountBS Bottle Sampler Manual
13
GB
Technical specifications
Feature
Specification
Principle of operation
Laser diode optical detection of actual particulates
Dimensions (typical)
H 530mm x W 210mm x D 410mm
Mass
Approx 18kg
Unit environmental working
temperature and humidity
+5°C to +40°C (–41°F to +104°F)
20–85% RH (tested at 30°C (86°F), non-condensing)
Unit environmental storage
temperature and humidity
+5°C to +40°C (–41°F to +104°F)
10–90% RH (tested at 30°C (86°F), non-condensing)
Moisture sensor calibration
±5% RH (over a compensated temperature range of +10ºC to +80ºC
(+50ºF to +176ºF)
Moisture sensor stability
±0.2% RH typical at 50% RH in one year
International codes
ISO 7 to 21, NAS 0 to 12
Contamination standards
Refer to Parker ‘Guide to Contamination Standards’ (DD0000015) on CD
ACFTD
: ISO 4406:1987; ISO 4406:1991; NAS 1638; AS4509E (Diff);
AS4509E (Cum); GOST 17216 : 2001
MTD
: ISO 4406:1999; NAS 1638; AS4509E (Diff); AS4509E (Cum)
Channel sizes
Channel Sizes: MTD µm(c)
ISO 4406:1999
NAS 1638
AS4059E (Diff)
AS4059E (Cum)
Jet Fuel
>4
4–6
4–6
>4
>4
>6
6–14
6–14
>6
>6
>14
14–21
14–21
>14
>14
>21
21–38
21–38
>21
>21
>38
38–70
38–70
>38
>25
>70
>70
>70
>70
>30
Channel Sizes: ACFTD µm
ISO 4406:1987
NAS 1638
AS4059E (Diff)
AS4059E (Cum)
GOST
17216:2001
>2
2–5
2–5
>2
>2–5
>5
5–15
5–15
>5
>5–10
>15
15–25
15–25
>15
>10–25
>25
25–50
25–50
>25
>25–50
>50
50–100
50–100
>50
>50–100
>100
>100
>100
>100
>100–200
Calibration
ACFTD
: fully traceable to gravimetric first principles
MTD
: via a certified primary ISO 11171 automatic particle detector using
ISO 11943 principles, with particle distribution reporting to ISO 4406:1996
Recalibration
Contact Parker Hannifin Manufacturing (UK) Ltd for advice
Fluid compatibility
Mineral-based oils and petroleum-based fuel. Contact Parker Hannifin
Manufacturing (UK) Ltd for advice on other fluid compatibility
Fluid management
Selectable on screen between 10 to 100ml
Viscosity range
1 to 300cSt