21
DB Series Nitrogen Generator
DB5-20 Model
Standby Settings – Standby Pressure
Accessed from settings menu. (press [F6])
• Use
UP
and
DOWN
keys to adjust Standby Pressure from
0 to 150 psi
• Default is 100 psi
• Press
[OK]
to store or
[ESC]
to exit without storing.
Standby Settings – Run Flow
Accessed from settings menu. (press [F6])
• Use
UP
and
DOWN
keys to adjust Run Flow
• Default is 60 scfh
• Press
[OK]
to store or
[ESC]
to exit without storing.
Standby Settings – Standby Flow
Accessed from settings menu. (press [F6])
• Use the
UP
and
DOWN
keys to adjust Standby Flow
• Default is 50 scfh
• Press
[OK]
to store or
[ESC]
to exit without storing.
4.8 Maintenance Timer Menu and Settings
Maintenance Timer
• Use
UP
and
DOWN
keys to scroll thru the menu items
- Set Timer
- Reset Timer
• Press
[OK]
to select or
[ESC]
to go back to main screen
Maintenance Timer – Timer Setting
• Timer results in MAINTENANCE TIMER alarm
• Use
UP
and
DOWN
keys to adjust maintenance timer in
Increments of 100 hours
• Default time is 1000 hours
• Press
[OK]
to store or
[ESC]
to exit without storing
Maintenance Timer – Reset Timer
• Use to reset maintenance timer after performing required
routine maintenance
• Press
[OK]
to RESET or
[ESC]
to exit without resetting
Содержание DB Series
Страница 1: ...DB Series Nitrogen Generator Owner s Manual Models DB5 20...
Страница 31: ...31 DB Series Nitrogen Generator DB5 20 Model 5 Select RTU Master Mode 6 Configure Network as follows...
Страница 41: ...41 DB Series Nitrogen Generator DB5 20 Model DB 5 10 PPM PCT FD 9 Flow Schematics...
Страница 42: ...42 DB Series Nitrogen Generator DB5 20 Model DB 15 20 PPM PCT FD...
Страница 55: ...55 DB Series Nitrogen Generator DB5 20 Model Notes...
Страница 56: ...56 DB Series Nitrogen Generator DB5 20 Model Notes...
Страница 57: ...57 DB Series Nitrogen Generator DB5 20 Model Notes...