background image

 

 

 

 

D I R E C T I O N S   F O R   U S E

 

NOTE:

   

These  directions  detail  a  basic  delivery  technique. 

However,  this  is  not  a  comprehensive  description  and  not  a 
substitute  for  a  training  course  that  emphasizes  a  practical, 
hands-on  approach  together  with  instruction  on  safe 
administration techniques.

 

 

1.  Maintain patient observation during 

procedure. 

2.  Turn ON unit by pushing in the ON / 

OFF switch. 

3.  Open N

2

O / O

2

 tank valves. 

4.  Using  Flow  Control  knob,  set  flow  rate  of  O

2

  to 

desired rate, keep bag about ¾’s full.   Rotate flow 
control knob upwards (clockwise) to increase flow.  
(See Figure 1, Item 4) 

Flow Control Knob 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5.  Set N

2

O concentration to desired level by rotating 

Concentration  Control  Knob  (See  Figure  1-Item 
5)  upwards  (counterclockwise)  to  increase 
concentration,  as  read  by  percentages  inscribed 
on  the  control  knob.    Rotate  slowly  until  desired 
level  is  achieved.    Practice  titration*  with  10% 
nitrous  upward  movements  every  60  seconds 
until  endpoint  achieved.    Patients  may  typically 
experience 

relief 

of 

anxiety, 

tingling 

in 

extremities,  and  euphoria.    Patients  typically 
require less than 50% nitrous. 

Concentration Control Knob     

   

 

 

 

 

 

 

 

6.  Flow Control knob may be re-adjusted to bring the 

total  flow  of  gases  back  to  desired  level,  when 
concentration  is  increased  or  decreased.    Total 
flow  is  equal  to  the  sum  of  right  and  left  tube 
readings.  Read  center  of  ball  float  on  flowmeter 
tube. 

 

 

 

7.  When  the  procedure  is  nearing  completion, 

amounts  of  N

2

O  should  be  decreased.  Terminate 

the  flow  of  N

2

O  and  deliver  100%  O

2

  to  begin  a 

minimum  postoxygenation  period  of  3  to  5 
minutes.  Assess  the  patient  for  appropriate 
recovery.  Administer  additional  O

2

  if  necessary. 

Titration  and  post-procedure  100%  O

2

  will 

minimize nitrous exposure to the patient treatment 
area,  potential  patient  side  effects  of  lethargy, 
headache,  or  nausea,  and  any  potential  adverse 
effects of nitrous diffusion into air filled cavities. 

8.  When procedure is finally completed, turn off 

both

 

control knob valves for gas shut off. 

9.  Place  the  ON  /  OFF  switch  (primary  shut-off 

mechanism) in the OFF position. (Push from back 
of ON / OFF switch.) NOTE:  If control valves are 
still open, gas flows should stop at this point. 

10.  Turn OFF the gas supply at the tank at the end of 

the day. 

 

BASIC DELIVERY TECHNIQUE: 

Practice  titration.    Titration  is  a  method  of 
administering  a  substance  by  adding  definitive 
amounts  of  a  drug  until  an  endpoint  is  reached.    For 
nitrous  oxide  /  oxygen  (N

2

O)  /  O

2

)  sedation,  N

2

O  is 

given in incremental doses until a patient has reached 
a comfortable relaxed state of sedation.  The ability to 
titrate  N

2

O  is a significant advantage because it limits 

the  amount  of  drug  to  that  which  is  required  by  the 
patient.    If  titration  is  done  properly,  the  patient  does 
not  receive  any  more  of  the  drug  than  is  necessary.  
The amount of N

2

O required by a patient on any given 

day or time varies. 
  
For  information  on  titration,  a  most  valuable  resource 
for  the  practitioner  is  the  Handbook  of  Nitrous  Oxide 
and Oxygen Sedation, written by Clark and Brunswick 
and published by Mosby (www.mosby.com).  This text 
is a concise and contemporary guide for nitrous oxide / 
oxygen administration. 

IMPORTANT NOTE: 

When  the  Concentration  Control  knob  is  open,  the 
Flow Control knob is closed, and there is no N

2

O flow 

indicated  in  the  flow  tube, the  Nitrous  Oxide Failsafe 
System will stop the flow of N

2

O. However, this safety 

feature  should  not  be  used  as  the  primary  shut  off 
mechanism.  The  control  knobs  are  for  primary  shut-
off. 

 

Setting 
Shown at 
60% 

Содержание C3000

Страница 1: ...Manufacturing Sarl Mailing Address Via privata Archimede 1 20094 Corsico Italy Phone 44 0 1271 313131 The Quality System for Porter Instrument is certified to ISO 13485 The scope of our registration i...

Страница 2: ...ection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire codes a...

Страница 3: ...s set to 0 5 Concentration Control Knob controls the concentration of N2 O but does not change the O2 flow rate 70 Maximum N2 O delivery Check Monthly 6 Emergency Air Valve automatically provides the...

Страница 4: ...itor O2 gas pressures at the beginning of each procedure to assure sufficient gas remains in the tank to complete the procedure SAFETY FEATURE CHECKS IMPORTANT These are safety features which you shou...

Страница 5: ...ing bag is attached to the metal portion of the flowmeter Bag Tee Installation to Flowmeter Screw knurled seal down tight onto flowmeter making sure the rubber washer is inside the seal nut When tight...

Страница 6: ...ation and post procedure 100 O2 will minimize nitrous exposure to the patient treatment area potential patient side effects of lethargy headache or nausea and any potential adverse effects of nitrous...

Страница 7: ...empty 1 Turn ON N2 O cylinder 2 Replace with full cylinder With N2 O concentration set both flows vary proportionally with no change in flow setting O2 regulator is varying pipeline pressure Check O2...

Страница 8: ...Similar to basic Model MXR design except different color coding for Oxygen white bands on flow tube and white buttons Flowmeter body is painted white and has a failsafe block cover that is also paint...

Отзывы: