INTENDED USE: 
Flowmeter for Analgesia 
 Delivery System 

This device is intended for patient use by an attending 
physician,  dentist  or  licensed  professional  properly 
trained in its use.  Porter Instrument recommends the 
user  be  thoroughly  familiar  with  the  use  of  Nitrous 
Oxide  -  Oxygen  Conscious  Sedation  for  patient 
analgesia  and  be properly trained in its administration 
prior  to  using  this  product.    For  training  requirements 
on  the  administration  of  Nitrous  Oxide  -  Oxygen 
Conscious 

Sedation, 

contact 

the 

appropriate 

regulatory authority in your country, state, or province.  
Training  is  recommended  to  provide  a  practical, 
hands-on  capability  and  an  understanding  of  the 
behavioral  aspects  of  Nitrous  Oxide  Sedation and will 
complement the safety features of this device. 

USE SCAVENGING 

Monitor for N

2

O in the patient treatment area to insure 

that  controls  are  effective  in  achieving  low  levels  of 
PPM (parts per million) exposure.  Contact your Porter 
dealer for details on monitors and testing. 
 

GAS SUPPLY CONNECTION 

After  installation  of  the  flowmeter,  connect  the  nitrous 
oxide and oxygen supply lines to the Diameter Indexed 
Safety  System  (DISS)  fittings  located  on  the  back  of 
the  flowmeter  unit.    It  is  important  that  the  regulators 
for both gases be set to give pressures in the range of 
53 PSIG ±2 PSIG. 
 
Confirm the absence of leaks at pressure connections 
on  the  unit.    Bubbles  will  appear  at  leaking  locations 
when a soap / water solution is used.  This procedure 
is recommended each time a cylinder is changed. 

MONTHLY  LEAK  CHECK

  (or  if  connections  to 

flowmeter  are  disconnected  and  then  reconnected, 
such  as  after  a  flowmeter  service  at  the  factory  or  if 
the  DISS/DISS  hose  is  replaced):    Leak  test  the 
flowmeter system for working pressure leaks.  After all 
hose  connections  are  tightened,  turn  both  control 
knobs to the off position and the on/off switch to the off 
position.  Confirm that the DISS Shut-Off Valves are in 
the open position.  Pressurize the sedation gas supply 
lines  with  50  PSI.    Observe  any  pressure  decay  after 
an overnight time period (5 PSI drop allowed

). 

 
Monitor  O

2

  gas  pressures  at  the  beginning  of  each 

procedure to assure sufficient gas remains in the tank 
to complete the procedure. 

 

 

 
 

 

SAFETY FEATURE CHECKS 

IMPORTANT

:

   

These  are  safety  features,  which 

you  should  routinely  check  to  assure  proper  function.  
If  any  of  these  safety  features  are  not  functioning 
properly,  contact  your  authorized  distributor  or  Porter 
Instrument  and  arrange  for  the  necessary  repairs.  
Porter  Instrument  recommends  the  repairs  be  made 
before reusing the device. 

 

NITROUS OXIDE FAILSAFE SYSTEM  

CHECK BEFORE EACH USE 

1.  Set the right-hand total Flow Control Knob to zero 

and  set  the  left-hand  Concentration  Control  Knob 
to  zero.    (See  Figure  1-Items  4  and  5:  Rotate 
control knobs to stop.) 

2.  Be  sure  O

2

  and  N

2

O  are  connected  to  your  MXR 

and  line  pressure  for  both  gases  is  50  PSIG 
(which is standard). 

3.  Turn  the  Concentration  Control  Knob  to  50%.  

There should be no flow of N

2

O.

  This is a check 

of the static position of the Nitrous Oxide Failsafe 
System valve. 
NOTE:    A  momentary  low  flow  of  N

2

O  (about  1 

L/min  for  about  a  second)  may  be  seen  if  the 
Concentration  Control  Knob  is  turned  to  zero 
before turning the Flow Control Knob to zero.  This 
is  N

2

O  gas  trapped  between  the  Nitrous  Oxide 

Failsafe  System  valve  and  the  Concentration 
Control  Knob  valve  and  is  a  normal  occurrence.  

No other N

2

O flow should be observed. 

4.  With  the  Concentration  Control  Knob  still  set  to 

50%, turn the Flow Control Knob to achieve a flow 
of  3  to  4  L/min  of  O

2

.    You  should  observe  an 

equal  amount  of  N

2

O  flowing  by  gradually  turning 

the  Flow  Control  Knob.    (Refer  to  Monthly 
Calibration Check on Page 4, to read ball float on 
flowmeter tubes.) 

5.  Interrupt  the  flow  of  O

2

.    This  will  check  the 

dynamic  status  of  the  Nitrous  Oxide  Failsafe 
System  valve.    This  can  be  done  by  either 
disconnecting  the  oxygen  hose  from  the  wall  or 
shutting  off  the  oxygen  at  the  tank.   

The  Nitrous 

Oxide  flow  should  drop  as  the  Oxygen  flow 
decreases,  stopping  completely  before  the 
Oxygen float drops to zero

   

W

ARNING

 

 
If the Nitrous Oxide Failsafe System fails to perform as 
indicated,  do  not  use  this  product  prior  to  repair.  
Improper  function  of  this  safety  feature  may  permit 
Nitrous Oxide to flow independently of the flow control 
knob,  potentially  allowing  Nitrous  Oxide  to  flow  to  the 
patient without Oxygen. 
 

 

 
 

Содержание C3000

Страница 1: ...Manufacturing Sarl Mailing Address Via privata Archimede 1 20094 Corsico Italy Phone 44 0 1271 313131 The Quality System for Porter Instrument is certified to ISO 13485 The scope of our registration i...

Страница 2: ...ection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire codes a...

Страница 3: ...s set to 0 5 Concentration Control Knob controls the concentration of N2 O but does not change the O2 flow rate 70 Maximum N2 O delivery Check Monthly 6 Emergency Air Valve automatically provides the...

Страница 4: ...itor O2 gas pressures at the beginning of each procedure to assure sufficient gas remains in the tank to complete the procedure SAFETY FEATURE CHECKS IMPORTANT These are safety features which you shou...

Страница 5: ...ing bag is attached to the metal portion of the flowmeter Bag Tee Installation to Flowmeter Screw knurled seal down tight onto flowmeter making sure the rubber washer is inside the seal nut When tight...

Страница 6: ...ation and post procedure 100 O2 will minimize nitrous exposure to the patient treatment area potential patient side effects of lethargy headache or nausea and any potential adverse effects of nitrous...

Страница 7: ...empty 1 Turn ON N2 O cylinder 2 Replace with full cylinder With N2 O concentration set both flows vary proportionally with no change in flow setting O2 regulator is varying pipeline pressure Check O2...

Страница 8: ...Similar to basic Model MXR design except different color coding for Oxygen white bands on flow tube and white buttons Flowmeter body is painted white and has a failsafe block cover that is also paint...

Отзывы: