CONTROLLING THE FLUID FLOW

The water source is selected using the two toggle 

switches on the front of the unit.

•  The left switch selects the fluid source between 

“External”, from the rear connector to the city water, 

or “Internal”, from the bottles inside the Integra.

•  The right switch selects the fluid source between the 

“Front” or the “Rear” bottle.

When using the internal bottle water source, room 

temperature, sterile, deionized or distilled water is 

preferred. 4 parts water may be mixed with 1 part  

FDA-approved mouthwash for a pleasant-tasting 

irrigation fluid. 

The fluid-control buttons on the touch panel control the 

output of the pump from the internal bottles only. If you 

are using an external water source, water flow  

is adjusted using the water control knob on the back  

of the unit.

If you are using the internal bottles as a fluid source, 

the flow is controlled by using the WATER (+) and 

(–) buttons on the touch panel at the front of the unit. 

When you press either button the display will start to 

flash to indicate that it is now showing the water flow 

rate (1 is the lowest; 7 is the highest). As you hold the 

button the flow will change as shown by the display. 

When you release the button the display will return to 

showing the power level after a few seconds. When 

in operation, the sound of the pump will speed up or 

slow down depending on the fluid flow setting. If fluid 

does not flow through the insert when the foot pedal is 

depressed, consult the Troubleshooting section later in 

these instructions.

HOW TO REPLACE THE FILTER DISK

If an external water source is used, the water filter disk 

on the rear hose of the unit should be replaced every 

3-4 months, or when it becomes dark or clogged with 

debris, or when water flow is impeded. Replacement 

filters are available from Parkell (Stock No. D419 – ten 

per package). To change the filter:

1.  Make sure all water valves from the water supply to 

the Integra are shut off before changing the filter.

2.  Disconnect the scaler water supply hose from the 

water supply at the quick-connect.

3.  Remove the old filter disk by holding it in one hand 

and twisting off the hose connectors on either side. 

They rotate in opposite directions.

4.  Install the new filter disk in the same position as 

the old one, and tighten the connectors by hand in 

reverse. Do not use a tool.

5.  Reconnect the scaler water supply hose to the water 

supply via the quick-connect, open the control valve 

and check for leaks.

6.  Note: If the filter becomes clogged and you do not 

have a replacement disk, the hose connectors may be 

locked together TEMPORARILY without the filter disk 

to allow emergency operation only.

ELECTRICAL CONNECTION

The scaler must only be powered via the Transformer 

Assembly that is supplied with the device by Parkell. The 

scaler and its separate transformer generate a minimal 

amount of heat. Avoid covering them, to allow normal 

cooling.

1.  Plug the transformer power cord connector into the 

“PWR” receptacle on the back of the scaler, with the 

arrow on the plug on top.

2.  Plug the foot pedal into the “FTPEDAL” receptacle  

on the back of the scaler, with the arrow on the plug 

on top.

3.  Plug the transformer power cord into a grounded 

electrical outlet.

USE OF AUTOCLAVABLE SHEATHS AND 

ULTRASONIC INSERTS

1.  The Integra handpiece is provided with the 

autoclavable handpiece sheath attached. Prior to 

use, the sheath should be removed and sterilized in a 

steam autoclave. To remove it, hold the handpiece at 

the base with one hand and turn the sheath counter-

clockwise with the other, taking care not to twist or 

damage the hose. The sheath will unscrew and slip 

off. Confirm that the orange O-ring is still present on 

the end of the handpiece.

2.  To replace the sheath, lubricate the O-ring on the 

handpiece with water to assure a good seal. Slide 

the sheath over the handpiece carefully to avoid 

damaging or dislodging the O-ring and gently thread 

the sheath onto the handpiece clockwise, until there is 

no gap between the two. Do not overtighten.

TURNING THE SCALER ON AND OFF

To turn the scaler on, press any button on the touch 

panel. The display will illuminate to indicate that the 

scaler is on.

To turn the scaler off, press the WATER minus (–) button 

and the POWER minus (–) buttons at the same time. The 

display will go out, indicating that the unit is off. The unit 

will remember any settings you made when it is turned 

on again.

4

External 

water 

source

Internal 

bottle

Rear  

bottle

Front 

bottle

Содержание INTEGRA D680

Страница 1: ...work properly with our scalers For more information contact Parkell WARNINGS The water supply to the scaler should always be turned off whenever the device is being connected disconnected or when not...

Страница 2: ...GS257 120V D403 GS438 230V 1 Instructions for Use INSERTS Magnetostrictive inserts for the Integra scaler are available separately and are not included with the basic scaler unit However they may be...

Страница 3: ...fluids other than water are used in the bottles the system should be flushed with plain water after each use To fill the bottles 1 Open the Integra cover You will see two bottles with the bottle lock...

Страница 4: ...connectors on either side They rotate in opposite directions 4 Install the new filter disk in the same position as the old one and tighten the connectors by hand in reverse Do not use a tool 5 Reconn...

Страница 5: ...e There is a small switch on the rear of the foot pedal to allow you to enable or disable the POWER button If the switch is in the UP position the POWER button is enabled If the switch is in the DOWN...

Страница 6: ...with an EPA registered low level HIV HBV claim to intermediate level tuberculocidal claim hospital disinfectant wiped of chemical residue with a water damp paper towel and dried Utilize the surface d...

Страница 7: ...FedEx and adequately insured Return all accessories with the unit and include a written explanation of the problem Pack in the original box add plenty of cushioning material and overbox the unit duri...

Страница 8: ...31 249 1242 E mail info parkell com www parkell com THIS PRECISION DENTAL DEVICE WAS DESIGNED MANUFACTURED AND IS SERVICED IN THE UNITED STATES OF AMERICA BY European Authorized Representative Not a d...

Отзывы: