parkell INTEGRA D680 Скачать руководство пользователя страница 3

I

 

 

INSTALLING THE SCALER

Locate the device where the control panel will be easy 

to reach during scaling procedures, and where the 

water filter at the rear of the unit may be periodically 

changed without difficulty. The scaler requires 

access to a grounded electrical outlet and an optional 

connection to a source of drinking-quality water from 

the building water supply. Alternatively, the unit may 

receive fluid exclusively from the internal bottle supply

WATER CONNECTION

If the clinician chooses to operate the Integra on city 

water, before plugging in the device, they should 

connect the scaler’s water line through the included 

filter disk to a drinking-quality water supply (15-35 psi 

optimal), free of sediment, at room temperature. An 

external, office-wide water filter is recommended to 

minimize frequent changes of the device’s in-line water 

filter. The male quick-connect that comes on the end 

of the water hose is the standard fitting presently used 

in the dental industry. If you are replacing a scaler that 

uses an Adec-type or other type of connection, remove 

the old fitting from the unit hose and attach it to the 

new water line. Check for leaks at all hose and filter 

connections before use, and tighten if necessary.

 

USING THE INTERNAL FLUID  

SUPPLY BOTTLES

The Parkell Integra is equipped with two internal 

bottles which can supply the fluid that flows through 

the handpiece for cooling and lavage during use. The 

bottles may be filled with drinking-quality water, or 

diluted with anti-microbials at a ratio of 4 parts water 

to 1 part chlorhexidine or mouthwash. If fluids other 

than water are used in the bottles, the system should 

be flushed with plain water after each use.

To fill the bottles:

1.  Open the Integra cover. You will see two bottles, with 

the bottle-lock holding them in place.

2.  Using finger and thumb, gently squeeze the two 

retaining lugs of the bottle-lock and remove it  

from the unit.

3.   Remove the bottles from the unit. Unscrew and 

remove the caps from them.

4.  Fill the bottles up to the fill line marked on the side.

5.  Replace the caps on the bottles, making sure they 

are screwed on all the way.

6.  Replace the bottles in the rear of the Integra, making 

sure they are standing upright in the base of the unit.

7.  Replace the bottle-lock over the bottles. Look for 

the word “Top” on the bottle-lock to make sure it 

is in correctly. As before, using finger and thumb, 

squeeze the two retaining lugs and  

gently press the bottle-lock down.  

Release the lugs. They should  

locate into the side of the clear  

cover and hold the bottles firmly  

in place.

8.  Completely close the Integra lid. If not  

fully closed, the unit will not be able to draw fluid 

from the bottles.

9.  Monitor the fluid level in the bottles regularly during 

use and refill as required. A constant fluid flow is 

necessary for correct operation of the Integra  

during scaling.

Foot Pedal

Power

External water  

flow regulator

External water 

connection

3

Bottle-Lock
(DO NOT DISCARD)

Apply lubricant  
to both ‘O’ rings.

The lid must be closed completely for the unit to function 
correctly with the internal fluid system. If the lid does not 
close completely or it sticks, then the two black ‘O’ rings 
on the underside of the cover may be dry and require a little  
lubrication. Apply a very thin layer of Vaseline to each ‘O’ 

ring with a Q-Tip. Just a thin layer is all that is necessary.

IMPORTANT:

Содержание INTEGRA D680

Страница 1: ...work properly with our scalers For more information contact Parkell WARNINGS The water supply to the scaler should always be turned off whenever the device is being connected disconnected or when not...

Страница 2: ...GS257 120V D403 GS438 230V 1 Instructions for Use INSERTS Magnetostrictive inserts for the Integra scaler are available separately and are not included with the basic scaler unit However they may be...

Страница 3: ...fluids other than water are used in the bottles the system should be flushed with plain water after each use To fill the bottles 1 Open the Integra cover You will see two bottles with the bottle lock...

Страница 4: ...connectors on either side They rotate in opposite directions 4 Install the new filter disk in the same position as the old one and tighten the connectors by hand in reverse Do not use a tool 5 Reconn...

Страница 5: ...e There is a small switch on the rear of the foot pedal to allow you to enable or disable the POWER button If the switch is in the UP position the POWER button is enabled If the switch is in the DOWN...

Страница 6: ...with an EPA registered low level HIV HBV claim to intermediate level tuberculocidal claim hospital disinfectant wiped of chemical residue with a water damp paper towel and dried Utilize the surface d...

Страница 7: ...FedEx and adequately insured Return all accessories with the unit and include a written explanation of the problem Pack in the original box add plenty of cushioning material and overbox the unit duri...

Страница 8: ...31 249 1242 E mail info parkell com www parkell com THIS PRECISION DENTAL DEVICE WAS DESIGNED MANUFACTURED AND IS SERVICED IN THE UNITED STATES OF AMERICA BY European Authorized Representative Not a d...

Отзывы: