DEVICE DESCRIPTION

The Parkell Integra™ is a magnetostrictive, auto-

tuning, ultrasonic tooth scaler that operates at either 

25 KHz or 30 KHz. It has an integrated, two-bottle, 

pump-driven fluid supply, and is also capable of 

obtaining fluid from an external water supply.  

The Integra automatically detects whether the 

handpiece contains a 25 KHz or 30 KHz insert, and 

switches to the correct operating frequency. The 

Integra has an expanded low-power range to improve 

patient comfort, and a power-boosting, user-controlled 

TURBO feature to increase scaling power when needed 

for heavy deposits.

INTENDED USE/INDICATIONS

For removal of calculus, plaque and oral debris from 

teeth during dental prophylaxis/scaling.

CONTRAINDICATIONS TO USE

Because of the potential for electromagnetic 

interference, this device should not be used on  

patients or by clinicians with cardiac pacemakers, 

internal defibrillators, intracorporeal fluid pumps or 

any other implantable electronic devices, or in close 

proximity to sensitive patient monitoring devices such 

as pulse oximeters.

IMPORTANT INFORMATION REGARD-

ING USE OF NON-PARKELL INSERTS:

Parkell ultrasonic scalers are compatible with most 

major brands of ultrasonic inserts. For optimal 

performance of your scaler and insert, we recommend 

using Parkell ultrasonic inserts. Parkell cannot 

guarantee that all non-Parkell inserts available on the 

market today or in the future will work properly with 

our scalers. For more information, contact Parkell.

WARNINGS

•  The water supply to the scaler should always be 

turned off whenever the device is being connected,

disconnected, or when not in use. 

•  Do not run this device dry. Water must always flow 

through tip during use to cool handpiece and insert. 

•  The power supply line cord functions as the AC 

power disconnect device for the scaler. 

•  The transformer should be located above and away 

from any sources of water. As with all electrical 

devices, the unit should not be immersed in water 

or other liquids. Do not reach for the device if it has 

fallen into liquid until power is disconnected, and 

do not use the device after it has fallen into liquid.

Return the device to Parkell for servicing. 

•  Modification of this device will void the warranty, and 

may violate safety codes, endangering the patient 

and/or the clinician.

•  This equipment is not suitable for use in the 

presence of a flammable anesthetic gas mixture 

(when used along with air or oxygen).

•  This equipment produces electromagnetic energy 

and may cause interference with other nearby 

electronic devices. Should this occur, changing the 

position or location of the device 

may be necessary. 

INSTRUCTIONS FOR USE 

STOCK NO. D680 

24-HOUR SAFETY HOTLINE

1-800-535-5053

CAUTION: RX ONLY 

This manual, in whole or in part, should not be considered a substitute for formal training in Ultrasonic Scaling.  

Appropriate professional education is STRONGLY RECOMMENDED prior to using this device clinically.

U LT R A S O N I C   S C A L E R

Shown with Parkell’s Burnett  

Power-Tip Insert (not included)

A0

0

545revC0317

Содержание INTEGRA D680

Страница 1: ...work properly with our scalers For more information contact Parkell WARNINGS The water supply to the scaler should always be turned off whenever the device is being connected disconnected or when not...

Страница 2: ...GS257 120V D403 GS438 230V 1 Instructions for Use INSERTS Magnetostrictive inserts for the Integra scaler are available separately and are not included with the basic scaler unit However they may be...

Страница 3: ...fluids other than water are used in the bottles the system should be flushed with plain water after each use To fill the bottles 1 Open the Integra cover You will see two bottles with the bottle lock...

Страница 4: ...connectors on either side They rotate in opposite directions 4 Install the new filter disk in the same position as the old one and tighten the connectors by hand in reverse Do not use a tool 5 Reconn...

Страница 5: ...e There is a small switch on the rear of the foot pedal to allow you to enable or disable the POWER button If the switch is in the UP position the POWER button is enabled If the switch is in the DOWN...

Страница 6: ...with an EPA registered low level HIV HBV claim to intermediate level tuberculocidal claim hospital disinfectant wiped of chemical residue with a water damp paper towel and dried Utilize the surface d...

Страница 7: ...FedEx and adequately insured Return all accessories with the unit and include a written explanation of the problem Pack in the original box add plenty of cushioning material and overbox the unit duri...

Страница 8: ...31 249 1242 E mail info parkell com www parkell com THIS PRECISION DENTAL DEVICE WAS DESIGNED MANUFACTURED AND IS SERVICED IN THE UNITED STATES OF AMERICA BY European Authorized Representative Not a d...

Отзывы: