Aim Laser on Center of Photodetector (Adjust Mirror 2) 

 
1) 

XEP software should already be open.

 

2) Check (A+B) voltage value (measures total intensity of light hitting the photodetector), which 
should be ~3-3.5 V if using ACTA tips provided by the lab. 

- The correct value will be different if you use your own tips.   

- If the (A+B) sum is significantly less than 3V, this indicates that either the ‘mirror 2’ 
alignment is way off, or the laser spot on the cantilever may actually be the reflected spot 
as mentioned above.  Adjust the ‘mirror 2’ knobs, first turning one knob until you find a 
local maximum of (A+B) voltage, then turning the other knob until you find the local 
maximum again.  Repeat as needed until the sum is at the appropriate value.   

- If after several tries, you still can’t get the sum high enough, remove the head and check 
mirror 2’s position visually – it should be roughly parallel to the base. Adjust mirror 2 
knobs as needed and then replace head 

3) Adjust ‘Mirror 2’ knobs to position laser spot on center of photodetector, reducing the (A-B) 
and (C-D) voltages to less than +/-0.3V. The XEP software will indicate the position of the laser 
as a red dot once it is reasonably near the center of the photodetector.  

Option A: Adjust the knobs while watching the laser position in the software and turning 
in the correct direction to move toward the center of the diagram/minimize the (A-B) and 
(C-D) voltages. 

Option B: Use the diagram below as a guide for which knob to turn and in which 
direction depending upon where the laser is shining.   

 
 
 
 

 


 

T = tighten 
L = loosen 

T1, L2 

L1, L2 

L1 

T1 

L2 

T2,L1 

T2, T1 

T2 

 

 
 
 

Mirror 2 Knobs 

Photodetector 

Содержание XE-100

Страница 1: ...ser will move on the quadrant photodetector This is relayed to a motor controlling the tip which adjusts the tip s vertical position until the laser is aimed in the center of the quadrant photodetecto...

Страница 2: ...ting tip and place it in the good tips box Ensure new tip is rested in place and that bearings fit into tip slots 4 Replace head and push flaps away from you until slightly tight This secures the head...

Страница 3: ...n Look at where the light from the microscope is shining on the tip assembly as this will help you to center the microscope Position laser on cantilever If you don t see the laser reflected off the ca...

Страница 4: ...e appropriate value If after several tries you still can t get the sum high enough remove the head and check mirror 2 s position visually it should be roughly parallel to the base Adjust mirror 2 knob...

Страница 5: ...o 1 00 Cantilever General 4 Click OK 5 Turn Head back on in software Frequency Sweep Window will open 6 The software will automatically detect the cantilever s resonance peak and select a driving freq...

Страница 6: ...B and C D are still less than 0 3 V 9 Shut acoustic enclosure 10 Click Approach button in Z stage control When the approach stops the light in the Z stage Control window will stop blinking Ensure tha...

Страница 7: ...will change the distance the tip hovers above the sample surface when scanning and can improve resolution of features surface tracking This can be done manually by changing the number in the Set Poin...

Страница 8: ...h XEI You can send images directly to the XEI data analysis software from the Buffer in XEP by right clicking on them Or you can open them from the spmdata or User Files folders Flattening your sample...

Страница 9: ...er as needed 5 Turn off laser switch unplug head and remove head 5a If the tip you used was good leave it on the AFM Otherwise remove head and place tip in the Bad Probes box 6 Remove sample from stag...

Страница 10: ...ion on center of photodetector A B 3 5V A B and C D 0 3V 6 Select appropriate mode scanner and frequency 7 Approach surface i Get optically close several mm ii Focus microscope 50 100um from surface m...

Отзывы: