MAGNAIR questions? Contact the Sunovion Customer Service line (1-888-394-7377).
7
Instructions for Use
For use with LONHALA
TM
(glycopyrrolate)
Inhalation Solution
Fig. C4-C5
Plug the AC adapter into the inlet
on the black battery door of
the Controller. Then plug the
AC adapter into the wall outlet.
C3. Close the battery door.
To close the battery door, push it closed until
you hear it Click into place.
See Fig. C3.
NOTE:
Rechargeable and disposable batteries
vary considerably in terms of storage life and
output. If used exclusively, batteries have an
operating life of ~2 weeks, based on high
quality disposable AA batteries that meet the
specifications listed in
Section H
and following
the cleaning procedure described in
Section F
.
To reduce the risk of battery leakage, it is
recommended to remove the batteries if you
plan to store the Controller for more than
30 consecutive days.
C4. Plug the AC adapter into the Controller.
To connect the AC adapter to the Controller,
place the Controller on a flat, stable surface.
The inlet is located on the underside of the
black battery door. Push the round end of
the AC adapter cord into the inlet.
Do Not
try
to insert the AC adapter into the blue colored
part of the Controller.
See Fig. C4-C5.
If you choose not to use the AC adapter,
it is strongly recommended to have an
extra set of batteries with you at all times.
STEPS FOR USING THE AC ADAPTER WITH MAGNAIR
The AC adapter will automatically adjust to the incoming voltage. The AC adapter will power
MAGNAIR with and without installed batteries.
C5. Plug the AC adapter into the wall outlet.
NOTE: the AC adapter will not charge the
batteries in the Controller.
See Fig. C4-C5.
C. STEPS FOR USING BATTERIES OR AC ADAPTER WITH MAGNAIR (cont.)
Black battery door
Inlet
AC adapter
Fig. C3
Close the battery door
of the Controller.