7.1 Nozzle blocked
Should the nozzle be blocked, you can clear it yourself as follows.
Figure
I
:
• Unscrew the nebuliser upper section (4) from the nebuliser lower
• Push the medication cup (5) up as far as the windows and then pull it out
of the nebuliser lower section.
• Press the nozzle (9) together with the airflow control (10) down and out of
the medication cup.
• Pull the airflow control up and off from the nozzle.
• Take the nozzle cleaner (7) out of the nebuliser lower section.
• Carefully poke through the three nozzle bore holes "A" to "C" (only from
below).
• Re-assemble the nebuliser (see Chapter 4.1).
7.2 Material resistance
Do not expose the nebuliser to temperatures above 137°C. In general, like all
plastic parts, nebulisers are susceptible to wear. Nebulisers can be sterilised
up to 300 times, but then they must be replaced. Even if the nebuliser has
been sterilised fewer than 300 times, it must still be replaced at least every 2
years. PARI GmbH recommends that certain components of the nebuliser
(nozzle, connection tubing etc.) be replaced every year. Note the following
when selecting cleaners/disinfectants:
- In general, the aldehydic group of cleaners/disinfectants is suitable for
cleaning and/or disinfecting the PARI LL nebuliser.
- The use of other cleaner or disinfectant groups has not been tested with
regard to the material resistance of this nebuliser.
- Only use an original PARI nozzle cleaner!
- Use the nozzle cleaner very carefully, because the small nozzle
bore holes can easily be damaged irreparably.
- If the nozzle becomes blocked repeatedly, replace it (part no.
019B2110).
7
Miscellaneous
Содержание LL
Страница 1: ...LL Vernebler 2010 PARI GmbH Spezialisten f r effektive Inhalation 012D0000 G 03 10...
Страница 3: ...1 A 3 11 2 4 5 6 9 10 13 8 7 12 B 3 1 C...
Страница 4: ...D H E I 4 5 10 9 C B A 7 8 F G...
Страница 26: ...22...
Страница 48: ...44...
Страница 71: ...67 el PARI PARI GmbH PARI PARI LL PARI PARI LL PARI PARI LL PARI TRACHEO SET 3...
Страница 72: ...68 4 1 A 10 9 5 8 4 3 1 2 LL 6 8 12 LL PARI CENTRAL 11 13 6 soft C 4...
Страница 73: ...69 el PARI 4 2 B 1 3 8 ml 6 1 2...
Страница 74: ...70 5 1 10 9 PARI PARI CENTRAL 5 2 D PARI CENTRAL E F G H 5...
Страница 75: ...71 el 6 1 PARI LL 6 4 1 6...
Страница 76: ...72 PARI GmbH 1 PARI GmbH PARI Year Packs 9 PARI 6 2 PARI LL A 11 13 1 3 2 4 8 5 9 10 6 4 soft PARI...
Страница 77: ...73 el 6 3 6 3 1 6 2 6 3 5 5 40 C 40 C 50 C 6 3 2 15 15...
Страница 78: ...74 6 3 3 4 6 3 4 M 6 3 5 PARI 041B4591 1 9 7 6 4 Robert Koch DGHM CE...
Страница 80: ...76 v 60 C 6 4 2 DIN EN ISO 11607 121 C 20 132 C 134 C 3 137 C DIN EN ISO 17665 DGKH 6 4 3 PARI 041G4550 1 9...
Страница 81: ...77 el 7 1 I 4 8 5 9 10 7 C 4 1 PARI 019B2110 7...
Страница 82: ...78 7 2 M 137 C 300 300 PARI GmbH 1 PARI LL 7 3 M 6 6 3 4 6 4 7 4 PARI LL...
Страница 85: ...81 el PARI BOY SX Year Pack 012G6300 PARI LL 012G6202 012B1800 012B1200 9...
Страница 86: ...82 012E1601 019B2110 022B0510 022E3050 LL 012E1024 PARI 1 2 m 041B4591...
Страница 87: ...83 el 2 1 m 041G4550 041E3500 soft 041G0740 soft Spiggy 041G0741 PARI SMARTMASK 45 041G0730 PARI 041G0500...
Страница 88: ...84 PARI PEP I 018G6100 PARI PEP II 018G6200 PARI PEP S 018G4000 PARI THERM 012G7110...
Страница 89: ...85 el T 49 0 8151 279 220 2010 PARI TRACHEO SET 025G0007...
Страница 90: ...86...
Страница 112: ...108...
Страница 154: ...150...
Страница 176: ...172...
Страница 198: ...194...
Страница 240: ...236...
Страница 281: ......