4.6 Visual inspection
Inspect all product components after each cleaning and disinfec-
tion. Replace any broken, misshapen or seriously discoloured
parts.
4.7 Drying and storage
• You can remove excess water more quickly by shaking all
parts. Remember to hold the cap of the nebuliser upper sec-
tion, otherwise the hinge may be damaged by shaking.
• Place all parts on a dry, clean and absorbent surface and let
them dry completely.
• For drying, between applications, and in particular during long
breaks, wrap the nebuliser in a clean, lint-free cloth (e.g. a tea-
towel) and keep in a dry, dust-free environment.
• During storage, make sure the nebuliser and accessories are
protected from continuous exposure to direct sunlight.
NOTICE
There is a
risk of damaging the nozzle bore holes irrepara-
bly
. Therefore, always use only an original PARI nozzle cleaner
and proceed extremely carefully. If the nozzle becomes blocked
repeatedly, replace the nozzle.