Manual cleaning
As an alternative to mechanical cleaning and disinfection, the
following manual method can be used:
EQUIPMENT
– Cleaning agent
– Brush
– Warm water
PROCEDURE
• Place all the parts in the cleaning solution.
• Clean the parts thoroughly with a brush.
• Thoroughly rinse the parts under hygienically tested water
and leave them to dry completely on a clean, dry and ab-
sorbent surface.
VALIDATION
The method has been validated using:
– A pH-neutral enzymatic cleaner, Korsolex
®
Endo Cleaner
0.5% (Bode)
– Approx. 50 °C warm water
– Duration of cleaning procedure: approx. 5 minutes
Manual disinfection
For manual disinfection, select a suitable instrument disinfect-
ant that has been proven to be effective and compatible with
the cleaning agent used.
EQUIPMENT
– Disinfectant
– Water
– 61 –