4.7 Storage
Store this product as described below:
• Wrap all individual components in a clean, lint-free cloth
(e.g., a tea towel).
• Store this product in a dry, dust-free place.
5
TROUBLESHOOTING
Fault
Possible cause
Remedy
No aerosol is
coming out of
the nebuliser.
The nebuliser
nozzle insert is
blocked.
Clean the nebuliser.
The connection
tubing is not con-
nected properly.
Check that the tubing connect-
ors are connected firmly to the
compressor and the nebuliser.
The connection
tubing is leaking.
Replace the connection tubing.
6
TECHNICAL DATA
6.1 General nebuliser data
Size
5
10 cm × 9.5 cm × 4.5 cm
Weight
5
27.5 g to 29.5 g
Operating gases
Air
Minimum compressor flow
3.0 l/min.
Minimum operating pressure
0.5 bar / 50 kPa
Maximum compressor flow
6.0 l/min.
Maximum operating pressure
2.0 bar / 200 kPa
Minimum fill volume
2 ml
Maximum fill volume
8 ml; PARI LC STAR: 6 ml
5) Without mouthpiece; unfilled.