Questions about Lamira?   Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) 

 

 
Step H5: Insert the Aerosol Head
  

Note: It is important that you do not touch the 

silver part of the Aerosol Head at any time 

during assembly. (H-5a) 
 
 
 
 
 
 
 
Check to make sure the Aerosol Head is labeled 

“For Amikacin Liposome Inhalation Suspension”.  

Do not use other eFlow Technology Aerosol 

Heads in the Lamira. (H-5b).   

 

Carefully grasp the Aerosol Head by the 2 flexible 

plastic tabs on each side of the Aerosol Head. 

Turn the Aerosol Head so that “For Amikacin 

Liposome Inhalation Suspension” is facing toward 

you and is at the top of the Aerosol Head (H-5b).  
 
 
 
 
 
Squeeze the two flexible plastic tabs together while 

inserting the metal arms and flexible plastic tabs 

into the Medication Reservoir (H-5c).  You should 

feel the flexible plastic tabs “grab” as you insert it.  

 
 

 

 

 

Step H6: Close your Handset by pushing the Aerosol 

Chamber together with the Medication Reservoir 

until you hear a “snap” (H-6).  If you do not hear a 

snap, open the Handset and check that the Blue 

Valve is seated properly (See Step H-4). 

 

 

 
 

 

 

 

    

  Assembling your Lamira Handset - continued 

H-6

 

H-5a

 

H-5b

 

H-5c

 

MEDICATION RESERVOIR WITH 4 HOLES 

FLEXIBLE PLASTIC 

TABS 

Reference ID: 4327567

Содержание Lamira

Страница 1: ...Reference ID 4327567...

Страница 2: ...Lamira nebulizer Before using your Lamira nebulizer read and understand all the Instructions for Use and save them for future reference If you do not understand any part of these directions contact t...

Страница 3: ...ion clean and dry all parts of the Handset after each use Follow the instructions in Section E to maintain and clean the Lamira Cleaning the Handset properly will help prevent the Aerosol Head from cl...

Страница 4: ...a Handset to be used for 28 days until the next ARIKAYCE drug box arrives o 1 ARIKAYCE Quick Start Guide o 1 Instructions For Use Insert o 1 Full Prescribing Information Insert Step C2 Check your Lami...

Страница 5: ...recommended to remove the batteries to reduce the risk of battery leakage Using the A C power supply The A C Power Supply will automatically adjust to the incoming voltage and will power the Controlle...

Страница 6: ...r Mouthpiece Wipe with clean disposable paper towel 1 second per part Wipe to remove residual medicine and then dispose of paper towel in trash with solid waste Rinse Prior to first use and after each...

Страница 7: ...ached to the mouthpiece as shown in the picture Your Handset including the Aerosol Head is not sterile Contamination and moisture may cause the growth of bacteria and the Aerosol Head can be affected...

Страница 8: ...systems Use only plain dry paper towels or wipes Do not use towels or wipes that have any chemicals added to them such as alcohol lotion or baby wipes Be careful not to harm the parts Do not wipe the...

Страница 9: ...e from soap disinfect your Handset Step F10 Disinfect the Handset including the Aerosol Head after cleaning In addition to cleaning ARIKAYCE from your Handset you must also disinfect your Handset to r...

Страница 10: ...Section F Step G2 Disinfect your Handset at the end of every day by boiling in distilled water To disinfect with boiling water boil the Handset parts including the Aerosol Head in a clean pot of disti...

Страница 11: ...ater and dry them well H 2 Step H3 Open the Handset The Medication Reservoir and the Aerosol Chamber are attached using a hinge First gently pull up on the tab of the Medication Reservoir H 3 This wil...

Страница 12: ...e Aerosol Head so that For Amikacin Liposome Inhalation Suspension is facing toward you and is at the top of the Aerosol Head H 5b Squeeze the two flexible plastic tabs together while inserting the me...

Страница 13: ...Connection Cord by lining up the bottom of the Connector with the bottom of the Handset H 8a and pushing the Connection Cord upward against the rear underside of the Handset H 8b until you hear the pa...

Страница 14: ...me throughout and well mixed I 1 Step I2 Lift orange cap from vial and throw away dispose of the orange cap I 2 Step I3 Grip metal ring on top of the vial Pull it down gently I 3 until 1 side breaks a...

Страница 15: ...on the LCD display I 8a You will also hear 1 beep and the status light will turn green The Lamira is now On After treatment begins the treatment screen I 8b will replace the start screen I 8a and aero...

Страница 16: ...change from dose to dose and may become longer unless the cleaning and maintenance instructions are followed See Maintaining Your Lamira Step I10 At the end of your treatment the following will happen...

Страница 17: ...at the Controller has power Secondly make sure the Connection Cord is correctly attached Thirdly check to be sure that your Handset has been assembled properly Section H Lastly check to be sure that t...

Страница 18: ...tart Controller 6 Flashing Lamira beeps 1x then beeps high low with the LED flashing and the LCD display flashing this symbol No ARIKAYCE detected Add ARIKAYCE If you have already added ARIKAYCE gentl...

Страница 19: ...el gently tap Handset and restart Controller 16 The Lamira does not stop automatically after all ARIKAYCE is consumed Clean and disinfect the Aerosol Head If condition continues contact the Arikares S...

Страница 20: ...imulation of label claimc 52 9 46 4 59 4 MMADd by NGIe m 4 45 4 38 4 52 GSDf 1 59 1 59 1 60 Respiratory Dose by NGI 5 m mg 327 1 316 5 337 6 Respiratory Fraction by NGI 5 m of delivered dose 55 6 53 8...