TROUBLESHOOTING
You can correct malfunctions that occur
during daily operation yourself. The follow-
ing tables will help you to locate and correct
the causes of malfunction.
Display indicator
(flashes alternately)
Possible cause
Solution
Cable connection to the
controller or to the
nebuliser handset
interrupted.
Check cable
connections.
No medication added.
Add medication.
Inhalation interrupted.
Resume inhalation by
restarting the device
(press ON/OFF button).
The maximum operat-
ing time of 20 minutes
for one application has
been exceeded.
Resume inhalation by
restarting the device
(press ON/OFF button).
Battery charge
level 50%.
Have new batteries
ready.
Battery flat.
Insert new batteries or
operate with the power
adapter connected.