PROCEDURE
• Place all the parts in the disinfection solution.
• Allow the solution to take effect as indicated in the instruc-
tions for use of the disinfectant.
• Thoroughly rinse the parts under hygienically tested water
and leave them to dry completely on a clean, dry and ab-
sorbent surface.
VALIDATION
The method has been validated using an aldehyde-containing
disinfectant:
– Korsolex
®
Basic 4% (Bode)
– Treatment time: 15 minutes
Manual cleaning with disinfection
The following chemical process may also be used as an altern-
ative to the mechanical process:
EQUIPMENT
– Cleaning disinfectant (a combination product for cleaning
and disinfection in a single work step)
– Warm water
– Basin
PROCEDURE
• Prepare the cleaning and disin-
fectant solution in the basin in
accordance with the manufac-
turer's instructions. The quant-
ity of solution required de-
pends on the number of
nebulisers to be cleaned.
– 62 –
Содержание 023G1801
Страница 1: ...LC SPRINT XLent 2015 PARI GmbH Spezialisten f r effektive Inhalation 023D1030 B 10 15...
Страница 2: ...de Gebrauchsanweisung 3 en Instructions for use 37...
Страница 45: ...Attach the connection tubing to the nebuliser Using with a mouthpiece Fit the mouthpiece onto the nebuliser 45...
Страница 69: ......