Problem
Check
Corrective action
Refer-
ence
page
The handle cannot be operated.
The handle may have been operated
beyond the highest or lowest position.
Turn the handle in the direction
opposite to the direction the handle
was turned excessively.
22
When the knee section is laid
flat, the leg section does not
become flat.
Check if the leg section base stay is
inserted at position (A), the normal
position for the stay adjustment fitting.
Lift the leg section base slightly, and
then move the leg section base stay
adjutment fitting to position (A).
20
The stretcher cannot be
positioned correctly.
Check if there are any obstacles
impeding the movement.
Remove any obstacles.
-
The stretcher cannot be moved. Check if the casters are locked.
Unlock the casters.
25
The stretcher does not move in
the intended direction.
Check if the center wheel is ON.
Switch OFF the center wheel.
26
The transfer board/Side-rail
cannot be lowered.
Check if the safety lock/dubble lock is
ON.
Release the safety lock/dubble lock.
28 to 31
The transfer board rattles.
Check if the stopper receptacles is
worm.
Contact your distributor for repair.
28
-
The height cannot be adjusted.
Check if there is an obstacles under or
behind the stretcher.
Remove the obstacles or anything
hindering the height adjustment.
-
There is abnormal noise during
operation.
Check if abnormal noise generate from
the oil cylinder.
Raise the stretcher to highest position,
then depress Up pedal a few times.
23
Check if abnormal noise generate from
parts other than the oil cylinder.
Lubricate the part that the noise
generate, then contact your distributor
for repair.
Product ID label
50
■
Problems may not always indicate faults. Check the following once again before requesting
repair.
■
If the problem persists after taking corrective actions, stop using the stretcher immediately, and
contact your distributor for repair.
■
Stretcher damaged in earthquakes, fires, floods or other incidents must be inspected and
repaired by your distributor. Otherwise, there is a risk of an injury due to malfunctioning caused
by deformation of the stretcher.
●
When requesting repairs or contacting your distributor, please provide the number on the product
ID label.
■
Product ID label location