PARAMOUNT BED KK-728EG Скачать руководство пользователя страница 26

25

6

 PATIENT TRANSFER

1

 

2

3

4

To transfer a patient from the stretcher to a bed

1

  Depress the center wheel locking pedal of the 

stretcher to switch it OFF, place the stretcher by the 
bed, and lock the casters by pressing down the caster 
lock pedal (red) (see p. 10 to 11).

2

  Align the height of the stretcher to the height of the 

bed by operating the height adjusting handle and 
place the transfer board over the bed (see p. 12 
and 14).

NOTE!

z

z

Set the bed slightly lower than the 
stretcher for easier transfer.

 WARNING

z

z

When placing the stretcher by a bed, prevent the transfer board or transfer 
board receptacle of the stretcher from sitting on the side rail (foldable type) 
of the bed. Otherwise, the stretcher may tip over, resulting in an injury or the 
side rail of the bed or the transfer board may get damaged or deformed.

Transfer board

Transfer board

Transfer board

Transfer board receptacle

[When the transfer board is up]

[When the transfer board is down]

3

  Unfasten the restraint belts from the patient and place 

them in an area where they will not hinder the transfer 
procedure. Grab the handles of the transfer mattress 
and transfer the patient onto the bed by slowly pulling 
the mattress.

4

  After transferring the patient, 

1

 Fold the side part of 

the transfer mattress that is further from the stretcher 
on the down side. 

2

 Roll the patient over in the same 

direction. 

3

 Pull out the transfer mattress toward the 

stretcher.

NOTE!

z

z

Fold the handles under the transfer 
mattress completely. If a small part of 
a handle is protruding, it gets caught 
under the patient’s body and the 
transfer mattress cannot be pulled out 
smoothly.

5

  Lay the patient on their back.

6

  Place the transfer mattress on the stretcher, return the 

restraint belts to the original position, raise the transfer 
board, and unlock the casters to move the stretcher. 
When moving the stretcher, raise the center wheel 
locking pedal to switch it ON for more stable stretcher 
movement and easier control (see p. 11).

Transfer board

Restraint belt

Center wheel 
locking pedal

Height adjusting handle

Caster lock pedals

Handle Transfer mattress

Restraint belt

Transfer mattress

Pull out

Fold the side part on the down side.

Fold the 
handles 
completely.

Transfer mattress

Caster lock pedals

Center wheel locking pedal

Содержание KK-728EG

Страница 1: ...Serious incident means any incident that directly or indirectly led might have led or might lead to any of the following a the death of a patient user or other person b the temporary or permanent ser...

Страница 2: ...wheel system 11 3 Height adjustment 12 4 Raising the backrest 13 5 Transfer board 14 6 Restraint belt 16 7 Oxygen cylinder holder 17 8 IV Pole C57E 19 9 Accessory rail 21 10 Transfer mattress 21 6 PAT...

Страница 3: ...ction base Accessory rail Oxygen cylinder holder Caster lock pedals Tray Base frame Height adjusting handle Main frame Leg section base IV Pole attachment hole Transfer board Restraint belt Transfer m...

Страница 4: ...es IV Pole C57E Load capacity 1kg hook Oxygen cylinder holder Load capacity 15kg Tray Polypropylene PP Load capacity 10kg Hook 3K003355 4 pieces Load capacity 3kg piece Patient target groups Patient a...

Страница 5: ...n could cause death or serious injury broken bones pressure or paralysis CAUTION Indicates that improper handling by ignoring this indication could cause light personal injury bruises wounds or cuts o...

Страница 6: ...on the stretcher z z There is a risk of injury caused by falling or toppling from the stretcher In particular be careful with children Do not crawl under or place your head hands or legs under the st...

Страница 7: ...rely secure the patient if using it with the two transfer boards locked in the horizontal level position z z Otherwise the patient could lose balance and fall out of the stretcher or fall over resulti...

Страница 8: ...e stretcher Otherwise the stretcher may be damaged or deformed or even catch fire Use care when handling the gas spring z z Gas spring is used for the backrest lever If you attempt to disassemble or i...

Страница 9: ...her is not in use store it with the casters facing in z z Otherwise people may trip over a caster and fall down or get injured Do not place objects weighing 10kg or more on the tray z z The tray could...

Страница 10: ...ve amounts of water vapor or oil vapor z z Locations that tend to be exposed to high temperatures or humidity low temperatures or dry conditions Note The optimum ambient conditions for the stretcher a...

Страница 11: ...operating only one caster lock pedal Caster lock pedals are located at the four corners of the stretcher allowing them to be operated at each side of head and foot ends To lock the casters z z Depress...

Страница 12: ...stretcher may tip over resulting in an injury When the center wheel is OFF z z Depressing the center wheel locking pedal sets the center wheel to OFF to allow the stretcher to move in any direction in...

Страница 13: ...heavy when it is turned close to the limit Do not turn the handle over the operating range Turning the handle forcibly over the range may result in damage or malfunction z z Do not put a leg on the ha...

Страница 14: ...equipment Grip Grip Do not hold any other place besides the grip To raise z z Hold the grip of the back section base with one hand hold the backrest lever on the back section base with your other han...

Страница 15: ...ay fall off and get injured To lower the transfer board To raise the transfer board To raise the transfer board 1 Slowly lift the transfer board which has been lowered in the direction shown in the fi...

Страница 16: ...wly lift the transfer board to a horizontal level position See To raise the transfer board in p 14 2 Move the transfer board in the direction shown by the arrow while keeping it in a horizontal level...

Страница 17: ...ay fall off or the stretcher may tip over resulting in an injury z z After placing a patient on the base back section base and leg section base of the stretcher use the two pairs of restraint belts to...

Страница 18: ...nd flowmeter may be deformed resulting in injury 7 Oxygen cylinder holder The oxygen cylinder holder can contain a 500L 10cm to 11 5cm diameter oxygen cylinder The oxygen cylinder holder can be set ei...

Страница 19: ...therwise the oxygen cylinder holder and flowmeter may be deformed resulting in injury Rotation stopper fitting Stopper Attachment part for horizontal setting Boss Thumb screw knob for locking a cylind...

Страница 20: ...g the sliding pipe of the IV Pole adjust the sliding pipe length and then lock the sliding pipe by tightening the loosened thumb screw knob WARNING z z When loosening the thumb screw knob firmly grip...

Страница 21: ...sliding pipe is loosened the sliding pipe may unintentionally fall resulting in injury z z Before use make sure the thumb screw knob for locking the sliding pipe is firmly tightened The sliding pipe m...

Страница 22: ...the hook or accessory rail resulting in injury 10 Transfer mattress A patient can be transferred to from a bed without being lifted For the procedures see 6 PATIENT TRANSFER on p 22 to 25 The mattress...

Страница 23: ...y or the side rail of the bed or the transfer board may get damaged or deformed Transfer board Transfer board Transfer board Transfer board receptacle When the transfer board is up When the transfer b...

Страница 24: ...their hands together on their abdomen and pull out the side part of the transfer mattress that was folded in step 4 Grab handles of the transfer mattress handles on the stretcher side and pull the mat...

Страница 25: ...ck pedal green to unlock the casters move the stretcher slightly away from the bed and raise the transfer board see p 14 WARNING z z When operating the transfer board place the patient s hands togethe...

Страница 26: ...they will not hinder the transfer procedure Grab the handles of the transfer mattress and transfer the patient onto the bed by slowly pulling the mattress 4 After transferring the patient 1 Fold the s...

Страница 27: ...as a support to stand up z z Do not transport a patient with the transfer board locked in the horizontal level position z z When transporting a patient be sure to fasten the restraint belts z z Operat...

Страница 28: ...bnormal noise If water gets on them immediately wipe it off Maintenance and inspection items to be completed by the user of the product Main Unit The position operations using the height adjusting han...

Страница 29: ...nium chloride e g Osvan Up to 0 2 Benzethonium chloride e g Hiamine Up to 0 05 Chlorhexidine gluconate e g Hibitane Up to 0 05 Sodium hypochlorite e g Milton It is highly recommended that you thorough...

Страница 30: ...e center wheel is set to ON z z Turn OFF the center wheel 11 The transfer board cannot be lowered z z Check if the safety lock is ON z z Release the safety lock 14 15 The transfer board rattles z z Ch...

Страница 31: ...od Paramount Bed keeps replacement parts necessary to maintain stretcher performance in stock for at least 8 years after the discontinuation of production 3 Questions regarding after sales service Con...

Страница 32: ...Translation of the original Instruction Made by Torindo Co Ltd 2022 04...

Отзывы: