PB-MF-MAN-046-REV1.0
Hot plate
6
2 Open fuse-holder by pulling the
flange
with
a
regular
screwdriver.
Figure 2. Open fuse-holder door.
3 Extract fuse holder using the
screwdriver.
Figure 3. Extract fuse-holder.
4 Replace fuses if necessary. Insert fuses in the fuse-holder in the correct position.
CORRECT
INCORRECT
Figure 4. Fuses position.
5 Insert the fuse-holder again, positioning it according to the voltage in the AC net.
115V POSITON
230V POSITION
Figure 5 Fuse holder position.
6 If the fuses blow again, unplug the equipment and contact technical service.
WARNING
For electrical safety reasons, never open the equipment. The power
supply has dangerous voltage levels.
Содержание LE7406
Страница 22: ...PB MF MAN 046 REV1 0 Hot plate 21...