13
Remote alarm terminal
The alarm of this unit can be informed at a remote location from this unit by connecting the external alarm
device to the remote alarm terminals. For the type and behavior of remote alarm output, refer to page 47
~
48.
The terminal of the remote alarm is installed at the
right side of the unit (See the figure on the point).
The alarm is outputted from this terminal. Contact
capacity is DC 30 V, 2 A.
When Buzzer key is pressed, the behavior of the
remote alarm is showed in Table.1.
Note:
In the door alarm, the remote alarm does not
work. Refer to page 47.
Table 1 The behavior of the remote alarm when pressing Buzzer key
Remote Alarm setting
(Refer to page 28
~
30)
Connecting
terminal
Normal
condition
Abnormal condition
(Including in the cases of power outage and
of where the power plug is pulled out.)
When pressing Buzzer key
ON:
Non-interlock with Buzzer key
COM.-N.C.
Close
Open
Open (Maintain in abnormality)
COM.-N.O.
Open
Close
Close (Maintain in abnormality)
OFF:
Interlock with Buzzer key
COM.-N.C.
Close
Open
Close (Return to normal)
COM.-N.O.
Open
Close
Open (Return to normal)
Remote alarm terminal