Chapter 5. Diagnostics
98
Sentinel™ User’s Manual
5.4 Flowcell Problems
If preliminary troubleshooting using the
Error Codes
“Displaying Diagnostic Parameters” on
indicates a possible flowcell problem, proceed with this section.
Flowcell problems fall into two general categories:
• Gas problems
• Pipe problems.
Read the following sections carefully to determine if the problem is indeed related to the flowcell. If the instructions in
this section fail to resolve the problem, contact Panametrics for assistance.
5.4.1 Gas Problems
Most gas-related problems result from a failure to observe the
Sentinel Flow Measurement System
installation
instructions. Refer to Chapter 1,
Installation
, to correct any installation problems.
If the physical installation of the system meets the recommended specifications, it is possible that the gas itself may
be preventing accurate flow rate measurements. The gas being measured must meet the following requirements:
•
The gas must be homogeneous, single-phase and relatively clean.
Although a low level of entrained particles may have little effect on the operation of the
Sentinel
system,
excessive amounts of solid (smoke) or liquid (steam) particles will absorb or disperse the ultrasound signals.
This interference with the ultrasound transmissions through the gas will cause inaccurate flow rate
measurements. In addition, temperature gradients in the gas flow may result in erratic or inaccurate flow rate
readings.
•
The gas must not excessively attenuate ultrasound signals.
Some gases (e.g., high-purity carbon dioxide, hydrogen, nitrogen, etc.) readily absorb ultrasound energy. In
such cases, an
E1
error code message will appear on the meter display screen to indicate that the ultrasonic
signal strength is insufficient for reliable measurements.
•
The gas soundspeed must not vary excessively.
The Sentinel system will tolerate relatively large changes in the gas soundspeed, as may be caused by
variations in gas composition or temperature. However, such changes must occur slowly. Rapid fluctuations in
the gas soundspeed, to a value that is considerably different from that programmed into the
Sentinel
, will result
in erratic or inaccurate flow rate readings. Refer to
“Displaying Measurements” on page lxxix
, and make sure
that the appropriate soundspeed appears in the soundspeed display.
Содержание Sentinel
Страница 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents x Sentinel User s Manual...
Страница 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Страница 55: ...Sentinel User s Manual 55 Chapter 2 Initial Setup 2 6 1 1 General cont Figure 13 General Tab Display...
Страница 57: ...Sentinel User s Manual 57 Chapter 2 Initial Setup 2 6 1 2 Fluid cont Figure 14 Fluid Tab Display...
Страница 59: ...Sentinel User s Manual 59 Chapter 2 Initial Setup 2 6 1 3 Errors cont Figure 15 Errors Tab Display...
Страница 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Страница 90: ...Chapter 3 Operations Sentinel User s Manual...
Страница 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Страница 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Страница 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Страница 132: ...Appendix B Data Records 132 Sentinel User s Manual B 2 Diagnostic Parameters Date Description of Service Performed By...
Страница 135: ...Sentinel User s Manual 135 Appendix C Brazilian INMETRO Approval Appendix C Brazilian INMETRO Approval...
Страница 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Страница 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Страница 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Страница 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Страница 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Страница 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Страница 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Страница 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Страница 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Страница 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Страница 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Страница 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Страница 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Страница 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Страница 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Страница 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Страница 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Страница 158: ...Index 158 Sentinel User s Manual...
Страница 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Страница 162: ......
Страница 164: ......
Страница 165: ......