Chapter 3. Operation
70
Sentinel™ User’s Manual
3.6 Configuring and Calibrating the Analog Outputs
Every Sentinel flow meter includes two built-in analog outputs (A and B) at terminal block J1. Before beginning
calibration of these outputs, an ammeter must be connected to the desired analog output. Both the zero-point and
the full-scale values for all of these outputs must be calibrated. After calibrating the outputs, which have a resolution
of
5.0 uA (0.03% of full scale), their linearity should be tested.
Note:
The zero point of the analog output may be set to either 0 mA or 4 mA. However, the calibration always uses
the 4 mA point, and the meter extrapolates this value to obtain the 0 mA point.
3.6.1
Configuring the Analog Outputs
Prepare for calibration by inserting an ammeter in series with the load on the desired output. See Figure 3-7 below to
identify the OUT(+) and RTN(-) pins.
Figure 7: Ammeter Connection (Output A)
Use the PanaView
Recorder Properties
to calibrate an Analog Output. To access these menu options, complete the
following steps:
1.
Make sure the system is connected to your computer, the system and the computer are turned ON, and
PanaView is up and running.
2.
Under
File
open the New Meter Browser.
3.
Expand the drop-down menus until you access the meter and its submenus. Expand the last meter submenu,
Unassigned I/O
.
4.
Double-click on
Recorder 1
or
Recorder 2
to bring up the Analog Output properties window. Under the
General
tab,
select the desired option from the
Type
drop-down menu (see
).
CAUTION!
The load on the analog output must not exceed 600 W. Do not connect the ammeter
directly across the analog output terminals.
B
RTN
SIG
B
RTN
A
SIG
A
1
4
Ammeter
–
+
Load
Terminal Block J1
Содержание Sentinel
Страница 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents x Sentinel User s Manual...
Страница 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Страница 55: ...Sentinel User s Manual 55 Chapter 2 Initial Setup 2 6 1 1 General cont Figure 13 General Tab Display...
Страница 57: ...Sentinel User s Manual 57 Chapter 2 Initial Setup 2 6 1 2 Fluid cont Figure 14 Fluid Tab Display...
Страница 59: ...Sentinel User s Manual 59 Chapter 2 Initial Setup 2 6 1 3 Errors cont Figure 15 Errors Tab Display...
Страница 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Страница 90: ...Chapter 3 Operations Sentinel User s Manual...
Страница 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Страница 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Страница 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Страница 132: ...Appendix B Data Records 132 Sentinel User s Manual B 2 Diagnostic Parameters Date Description of Service Performed By...
Страница 135: ...Sentinel User s Manual 135 Appendix C Brazilian INMETRO Approval Appendix C Brazilian INMETRO Approval...
Страница 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Страница 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Страница 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Страница 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Страница 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Страница 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Страница 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Страница 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Страница 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Страница 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Страница 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Страница 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Страница 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Страница 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Страница 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Страница 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Страница 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Страница 158: ...Index 158 Sentinel User s Manual...
Страница 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Страница 162: ......
Страница 164: ......
Страница 165: ......