Installing Panametrics Measurement & Control Ultrasonic Flow
Transmitters in a Potentially Hazardous Area
When installing this apparatus in a potentially hazardous area, the following requirements must be met:
•
The installation shall comply with IEC 60079-14.
•
Apparatus Markings:
Ex d IIC T6 Gb
•
Supply connection wiring shall be rated at least 10°C above rated ambient temperature of 60°C. Cable entries
are 3/4” NPT.
•
Certificate Number: IECExFMG12.0011X
Complies with IEC 60079-0: 2011 and IEC 60079-1: 2007
Manufacturer: Panametrics Sensing, Billerica, MA.
•
The certification covers the following Panametrics product types — XMT868, XMT868i, XGF868, XGF868i, XGN868,
XGN868i, XGS868, XGS868i, XGM868, XGM868i, IGM878 &: IGM878i.
•
This apparatus is equipment category 2 and may be employed in Zones 1 and 2. It may not be employed in a
Zone 0 area. This apparatus is certified for gas hazards only.
•
Enclosure Material: Si=95, Mg=0.35%, Mn=0.5%, Al = remainder.
•
This apparatus must be securely mounted using the mounting fixtures provided on the enclosure.
•
All covers are fitted with a locking device. This device must be tightened after the cover is replaced so as to
preserve the protection offered by the enclosure.
•
Modifications to the flameproof enclosure are not permitted.
•
Cable glands of an approved flameproof design are required. These must be installed according to the
manufacturer's instructions. Where the cable glands are provided by Panametrics Sensing, the manufacturer’s
instructions, as supplied to Panametrics Sensing, will be included in the documentation.
•
Connecting cable shall be mounted securely and protected from mechanical damage, pulling and twisting.
•
Unused entries must be sealed using a certified threaded plug.
•
The apparatus should be de-energized before opening.
•
If live working is necessary, the apparatus may only be opened when the surrounding atmosphere is known to be
safe. Measures must be taken to ensure that it remains safe while the unit is open, such as a system of hot work
permits.
•
The apparatus uses standard printed circuit assemblies. Always quote the apparatus serial number when
ordering spares.
•
This flameproof apparatus is only designed to be installed in a potentially hazardous area. This apparatus has
not been designed for use in a safe area, where it would be subject to CD 2006/95/EC (LVD). For a safe area
installation, an external mains disconnect must be provided close to this apparatus. If in doubt, consult
Panametrics.
Содержание DigitalFlow XGF868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 7: ...DigitalFlow XGF868i Startup Guide v Preface...
Страница 8: ...Preface vi DigitalFlow XGF868i Startup Guide no content intended for this page...
Страница 34: ...Chapter 1 Installation DigitalFlow XGF868i Startup Guide 24...
Страница 46: ...Chapter 2 Initial Setup 36 DigitalFlow XGF868i Startup Guide...
Страница 48: ...Chapter 2 Initial Setup DigitalFlow XGF868i Startup Guide 38...
Страница 68: ...Appendix A CE Mark Compliance 58 DigitalFlow XGF868i Startup Guide...
Страница 74: ...Appendix B Data Records 64 DigitalFlow XGF868i Startup Guide...
Страница 80: ...Index 4 DigitalFlow XGF868i Startup Guide...
Страница 82: ...Warranty 6 DigitalFlow XGF868i Startup Guide...
Страница 84: ......
Страница 86: ......
Страница 88: ......
Страница 89: ......