DigitalFlow™ GC868 Service Manual (1 and 2-Channel)
3-2
Chapter 3. Diagnostics
Table 3-1 below lists the available diagnostic parameters for the Model GC868. The
first column in the table shows the parameter as it appears on the option bar, while
the second column shows the parameter as it appears in the prompt area after it
has been selected.
Table 3-1: Available Diagnostic Parameters
Option
Bar
Screen Display
Description
Good
Bad
SS up
UP Sig Strength
Displays the signal strength for
the upstream transducer.
58–68
<58 or >68
SS do
DN Sig Strength
Displays the signal strength for
the downstream transducer.
58–68
<58 or >68
SNDSP
Soundspeed Ft/s
Displays the measured sound
speed of the fluid.
N.A.
N.A.
Tup
UP Transit S usec
Displays upstream transit time of
the ultrasonic signal in
microseconds.
N.A.
N.A.
Tdown
DN Transit usec
Displays downstream transit time
of ultrasonic signal in
microseconds.
N.A.
N.A.
DELTA
DeltaT usec
Displays the difference in µsec
between upstream and
downstream transit times.
N.A.
>30 nsec
REYN#
Reynolds #
Displays the Reynolds number.
<100,000,00
0
N.A.
Tot K
K(Re)
MultiK*KFactor
Displays the combination of all K
factors.
N.A.
N.A.
PEAK%
PEAK%
Displays the percentage of peak
(set to +80 by default).
N.A.
N.A.
Qup
UP Signal Q
Displays the signal quality for the
upstream transducer.
Š 1,200
–400 to +400
Qdown
DN Signal Q
Displays the signal quality for the
downstream transducer.
Š 1,200
–400 to +400
AMPup
UP Amp Discrim
Displays the value for the
upstream transducer amplitude
discriminator.
24 ± 5
<19 or >29
AMPdn
DN Amp Discrim
Displays the value for the
downstream amplitude
discriminator.
24 ± 5
<19 or >29
CNTup
UP DAC Counts
Displays the AGC DAC count for
upstream gain setting.
N.A.
N.A.
CNTdn
DN DAC Counts
Displays the AGC DAC count for
downstream gain setting.
N.A.
N.A.
P#up
UP +- Peak
Displays signal peaks for the
upstream transducer.
100-2,300
<100 or >2,300
P#dn
DN +- Peak
Displays signal peaks for the
downstream transducer.
100-2,300
<100 or >2,300
TEMP
Temperature deg F Displays the gas temperature
(0/4-20 mA or RTD input).
N.A.
N.A.
PRES
Pressure psia
Displays the gas pressure
(0/4-20 mA or RTD input).
N.A.
N.A.
Содержание DigitalFlow GC868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Warranty iv DigitalFlow GC868 Service Manual 1 and 2 Channel no content intended for this page...
Страница 20: ...Chapter 1 Calibration 1 11 DigitalFlow GC868 Service Manual 1 and 2 Channel no content intended for this page...
Страница 22: ...Chapter 1 Calibration DigitalFlow GC868 Service Manual 1 and 2 Channel 1 13 no content intended for this page...