DigitalFlow™ GC868 Service Manual (1 and 2-Channel)
2-4
Chapter 2. Error Codes and Screen Messages
2.17 E15: Equation
Limit
Problem:
The steam measurement may not be within 1% accuracy.
Cause:
The temperature or pressure reading is outside the accuracy
specification of the GC868 steam equations.
Action:
Adjust the pressure reading to within the range of 139.9685 kPa to
2242.9435 kPa (20.8717 to 326.4482 psi), or the temperature reading to
within the range of 373.1500°K to 810.9278°K (100 to 212°F).
2.18 E16: High Signal
Problem:
The meter signal strength exceeds the limits entered via the User
Program.
Cause:
A signal that exceeds the programmed limits is probably caused by the
entry of an improper value in the TRANSDUCER or PIPE submenu of the
User Program.
Action:
Check the values programmed into the TRANSDUCER and PIPE
submenus, as described in the Startup Guide and the Programming
Manual. Bypass or remove the preamp.
2.19 Screen
Messages
A variety of messages may appear on the display screen during the performance of
a task. Since the error codes have already been discussed in this chapter and the
locator bar messages are discussed in detail in Chapter 3,
Operation
, of the
Startup
Guide
, they will not be repeated here. All other messages are listed in Table 2-1
below.
Table 2-1: Screen Messages
Message
Meaning
Power Up Messages
Backup Battery FAIL
The backup battery that powers the non-volatile RAM has failed.
Contact the factory.
Backup Battery OK
The backup battery that powers the non-volatile RAM has passed.
EPROM sum = XXXX
Record the EPROM sum at the initial power up and periodically
thereafter.
FRIGID_INIT Executed
The NVR (non-volatile RAM) was automatically initialized due to a
memory fault. Contact the factory.
NVR FAIL
The non-volatile RAM has failed. Contact the factory.
NVR OK
The non-volatile RAM has passed.
Measurement Mode Messages
All Logs Cleared! hit key
This message may appear during one of the following tasks:
1. clearing a log - there are no more logs to clear
2. logging data - user has hit
[ENT]
instead of selecting an old log
or entering a new name
3. printing a log - there are no logs to print
4. displaying a log - there are no logs to display
Hit any key to resume taking measurements.
All Sites Cleared! hit key
This message may appear during one of the following tasks:
1. saving a new site - a new site name was not entered or overwrite
existing site was not selected
2. recalling a site - there are no site files to recall
3. clearing a site - there are no sites to delete
4. printing a site - there are no sites to print
5. displaying a site - there are no sites to display
Hit any key to resume taking measurements.
Do you want to SAVE?
This message appears upon exiting the
User Program
, if the new site
data has not been saved. Failure to save will result in loss of the new
data next time the site data is changed or recalled.
Содержание DigitalFlow GC868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Warranty iv DigitalFlow GC868 Service Manual 1 and 2 Channel no content intended for this page...
Страница 20: ...Chapter 1 Calibration 1 11 DigitalFlow GC868 Service Manual 1 and 2 Channel no content intended for this page...
Страница 22: ...Chapter 1 Calibration DigitalFlow GC868 Service Manual 1 and 2 Channel 1 13 no content intended for this page...