60
Owner’s Manual
Ashwood,E.R.(ed.),Tietz Textbook of Clinical
Chemistry. Philadelphia:W.B. Saunders Company
(1994),959.
Performance Characteristics
• Precision:
Standard deviation (SD) for each glucose
concentration < 100 mg/dL (5.55 mmol/L) and
coefficient of variation (CV) for each glucose
concentration ≥ 100 mg/dL (5.55 mmol/L)
is ≤ 5.0 mg/dL (0.278 mmol/L) and ≤ 5.0%,
respectively.
Intermediate precision
Control Solution Level
(mmol/L)
Low
(1.7~2.8)
Normal
(5.3~8.0)
High
(15.6~23.3)
Pooled Mean (
mmol/L
)
2.5
5.6
19.6
SD
2.8
3.1
8.6
CV (%)
-
3.1%
2.4%
Repeatability
Blood Glucose
(mmol/L)
1.7~
2.8
2.9~
6.1
6.2~
8.3
8.4~
13.9
14.0~
22.2
Pooled Mean
(
mmol/L
)
2.2
4.5
7.2
11.0
18.0
SD
2.7
3.5
4.6
6.5
8.3
CV (%) 6.9% 4.3% 3.5% 3.3% 2.6%