Owner’s Manual
Probable Cause Actions
Error in performing
Read the instructions
the test.
thoroughly and repeat
the test again.
Control solution bottle Shake the control
not shaken well.
solution bottle vigorously
and repeat the test again.
Expired or
Check the expiration
contaminated
date and the discard date
control solution.
of the control solution.
Control solution that
Control solution
is too warm or too cold. should come
to room temperature
(between 39°F/4°C and
86°F/30°C) before
testing.
Test strip deterioration. Please repeat the test
with a new test strip.
v
The control solution test is out of
range.