The Filter Capsule is leaking where it joins  
the Docking Station 

 

Ensure that the o’ring seal on the Docking Station 
is in position and intact. Ensure that Filter Capsule 
is fully engaged in the Docking Station. To confirm 
engagement, try to rotate the Filter Capsule to the 
left after connection. It should not rotate if fully 
engaged. 

The Filter Capsule does not disconnect from  
the Docking Station  

Ensure the release button on the Docking Station 
has been fully lifted. If disconnecting from a QPoint 
Docking Station - Tap Assembly with Quick Connect 
Inlet, the Docking Station will need to be held to 
avoid rotation. Grasp the Filter Capsule and rotate 
left to disengage. 

Water flow through the Filter Capsule is slow  

If the Filter Capsule has been disconnected, air may 
have entered and caused air locking which can 
significantly reduce the flow of water through the 
device. If this is the case, the Filter Capsule should 
be replaced with a new one. If the Filter Capsule has 
not been disconnected then it is possible that the 
reduction in water flow may be as a result of heavy 
contamination of the water supply with particles, rust 
or micro-organisms. This particulate contamination 
may be intermittent and/or related to maintenance 
or building work associated with the water system. 
The Filter Capsule should be replaced with a new 
one when the flow rate becomes unacceptable. If 
appropriate contact your local Pall representative or 
distributor for suitable pre-filtration advice. 

The Filter Capsule has been replaced, but  
the water flow is still reduced 

Ensure that the Filter Capsule is fully engaged with 
the Docking Station. Flow rates through the Filter 
Capsule are dependent on water pressure. If your 
water pressure is less than 1 bar, the pressure may 
not be enough to achieve sufficient flow through the 
device. Please contact a plumber or local estates 
department to advise if the water pressure can be 
increased. 

The Filter Capsule has been dropped on  
the floor. Can it still be used?  

No, you will need to replace the Filter Capsule.

Upstream, non-filtered water samples are 
needed for water quality testing 

 

If a water sample is required for testing between 
Filter Capsule exchanges, remove the existing 
Filter Capsule and dispose of according to local 
procedures. To achieve water flow from the water 
outlet, disconnect the Docking Station. Collect the 
water sample(s) as per local guidelines. Immediately 
reconnect the Docking Station and install a new  
Filter Capsule. 

If I have a problem who should I contact?

 

If you have a problem which you are unable to 
resolve, please see 

www.pall.com/QPoint

, and/

or contact your local Pall representative, distributor 
or retailer to discuss and if appropriate to return 
product(s) for further investigation. Products for  
return must be placed in a re-sealable bag and 
returned following instructions given by your local  
Pall representative, distributor or retailer.