Assemble System
Installation
1. Verify all components have been provided. (See
Tray Assembly and Fittings/Part Bag figures on
reverse side)
2. Cut PharMed tubing (yellow) to lengths as specified
(See page 3)
3. Slide tube clamps onto tubing and insert fittings into
tubing as shown on back page. Slide clamps over
fitting ends and pinch together (use pliers if
necessary to assist)
4. Lay drip tray on work counter. (Note: if there is any
rocking of tray, the bottom feet can be adjusted by
rotating as necessary as shown in figure 1)
5. Slide stir plate into left compartment of drip tray
underneath reservoir bracket. Attach capsule stand on
raised boss in center of drip tray with screw in threaded
hole. Turn screw until snug fit, but allows the stand to
move. (See figure 2)
6. Connect 3-way valves to fittings on reservoir outlet
and inlet. Attach male luer plugs onto female luer
connectors on 3-way valves. (See figure 3)
7. Place reservoir onto stir plate with outlet toward rear
of plate. (See figure 4)
8. Connect “long tube” to 3-way valve at reservoir outlet
and to fitting on rear gauge mount. (See figure 4)
9. Attach pump head onto pump with supplied screws.
(See figure 5)
Figure 2
Figure 3
Figure 4
Figure 1
Stir plate
Capsule
stand
3-Way valve
on inlet
3-Way valve
on outlet
Male luer
pugs
Long tube
Reservoir
3-way
valve
Pump
Pump
head
Mounting
screws
Figure 5