7.
Fluid Addition
WARNING!
Power must be ON to allow the UCI to regulate pressure
Once the single-use system is inflated, the system can be filled with fluid. During fluid addition the
system will regulate the pressure in the single-use system by allowing gas to exhaust from the rear
of the UCI.
5. Complete the sterile connection by pushing the
male connector’s plunger into the female connector.
6. Engage the sensor into the Allegro single-use mixer
by pushing the sensor into the mixer biocontainer,
collapsing the bellows.
7. Engage the end of the bellows in to the sensor
support bracket to maintain optimum sensor
position in the mixer.
1. Connect a suitable fluid addition line to the mixer
inlet line and add fluid
WARNING!
Filling phase must be observed to ensure pressure
does not exceed specifications as outlined below
Approximate fill volume can be observed using the
volume indicator on the inside of the tote (viewed
through the door during filling)
If the pressure continues to rise above 20 mbar on the
UCI during filling, reduce the fluid filling rate. The fill rate
for the 50 L mixer should not exceed 10 L/min