Calibrating to 100% Saturation will also calibrate
mg/L corresponding to 100% Saturation.
Calibrate 100% Saturation:
1
Rinse the probe with DI water or rinse solution
and let it dry. Do not touch the membrane.
2
Press MODE to select % Saturation mode.
3
Hold the probe in the air with the sensor
downwards. Let the reading stabilise.
4
Press CAL. The ‘CAL’ indicator blinks.
5
Press HOLD/ENTER to confirm. ‘CO’ will be
displayed.
Calibrate 0% Saturation:
1
Rinse the probe with DI water or rinse solution
and let it dry. Do not touch the membrane.
2
Press MODE to select % Saturation mode.
3
Dip the probe into the zero oxygen solution. Let
the reading stabilise.
4
Press CAL. The ‘CAL’ indicator blinks.
5
Press HOLD/ENTER to confirm. ‘CO’ will be
displayed.