2 Rinse the probe thoroughly with
deionised water, and then rinse
with a small amount of calibration
standard.
3 Dip the probe into the calibration
standard. Immerse the probe tip
beyond the upper steel band. Stir
the probe gently to create a
homogeneous sample.
4 Wait for the measured conductivity
value to stabilize.
5 Press CAL/MEAS to enter
conductivity. The CAL indicator
will appear in the upper right
corner of the display.
6 Press the or keys to change
the value on the primary display to
match the value of the calibration
standard.
7 Press ENTER to confirm
calibration value. The meter
returns to the MEAS
(measurement) mode.