3.3.2
Bottom
view
of
the
sensor
The
tube
of
the
pump
(plug
connection)
is
plugged
into
the
correspondent
socket
at
the
bottom
side
of
the
sensor.
3.4
Cleaning
the
sensor
3.4.1
How
to
clean
the
sensor
First,
remove
the
pump
tube
(plug
connection)
from
the
sensor
housing
at
the
bottom
side
of
the
sensor
(see
picture
above).
Then,
the
two
hexagon
cap
screws
have
to
be
loosened.
To
do
so,
use
the
correspondent
wrench.
After
that,
the
measuring
volume
can
be
taken
away
from
the
sensor.
Move
the
aerosol
guide
tube
from
the
bottom
side
at
the
connection
of
the
pump
tube
slowly
upwards.
Pump
tube
with
connection
for
sensor
Connection
of
the
pump
tube
to
the
bottom
side
of
the
sensor
Open
the
two
hexagon
screws
at
upper
side
of
the
sensor