5
Operation and settings
5200-en_V2.0_06/20
53
5.2.12 Weather station calibration screen
The following weather station sensors can be calibrated in the weather sta-
tion screen.
• Temperature sensor
• Air pressure sensor
• Humidity sensor
Calibration takes place by comparing the measured values to the measured
values of a transfer standard.
These types of calibration are possible:
• One-point adjustment (normally under field conditions):
Determination of the slope factor (scale); offset remains 0.
• Two-point or multiple-point adjustment (e.g. calibration lab):
Determination of the slope factor (scale) and the offset applying regres-
sion calculation.
Menu path
Main menu > accessories > calibrate weather station
Weather station calibration screen
Note
Changes are not applied until after the system is restarted.