Letting device cool off
The drying section IADS and the sampling head get hot during operation.
After switching off the device, allow it to cool off sufficiently before touching
interior components. How long it takes to cool off is a factor of the ambient
temperature on site.
CAUTION
Risk of burns from hot components
Components can remain hot even after the device is switched off. Before
touching components;
4
Allow the device to cool off sufficiently.
Removing weatherproof housing
AQ Guard Ambient – detaching wall bracket and sampling head
1.
Remove the three screws from the underside of the wall bracket.
ð
The device can now be detached from the wall bracket.
2.
Detach the device from the wall bracket.
3.
Unscrew the sampling head.
4.
Lift off the weatherproof housing.
5.
Screw the sampling head back onto the drying section. Verify that the
O-ring is in place.
Install the weatherproof housing in the reverse order of removal.