English
42
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
(according to EN 60601-1-2:2007; 5.2.2.1 Table 2)
MultiStim SWITCH is intended for use in the electromagnetic environment specifi ed below. The
customer or user of the MultiStim SWITCH should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic envi-
ronment – guidance
Electrostatic
discharge (ESD)
according to
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic
tile. If fl oors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Electrical fast tran-
sient / burst
according to
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Not applicable
Not applicable
Surges according to
IEC 61000-4-5
± 1kV line(s) to line(s)
± 2kV line(s) to earth
Not applicable
Not applicable
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines accord-
ing to
IEC 61000-4-11
< 5%
U
r
(>95% dip in
U
r
)
for 0,5 cycle
40%
U
r
(60% dip in
U
r
)
for 5 cycles
70%
U
r
(30% dip in
U
r
)
for 25 cycles
< 5%
U
r
(>95% dip in
U
r
)
for 5s
Not applicable
Not applicable
Power frequency
(50/60 Hz) mag-
netic fi eld accord-
ing to
IEC 61000-4-8
3 A/m
3 A/m
Power frequency mag-
netic fi elds should be
at levels characteristic
of a typical location in
a typical cornmercial or
hospital environment.
U
r
is the a.c mains voltage prior to application of the test level.
G
(
M
c
I
C
a
I
R
a
I
N
XS190174B MultiStim SWITCH GB.indd 42
XS190174B MultiStim SWITCH GB.indd 42
12.03.10 13:53
12.03.10 13:53