3 Technical description
The MultiStim SWITCH generates reproducible square pulses with adjustable frequency and continuo-
usly adjustable stimulation current. The adjustment range of the impulse current: 0.0 – 6.0 mA if stimu-
lation cannulae are being used, and 0,0 – 20 mA if stimulation catheters are being used.
The corresponding symbol will be shown synchronously with a short beeping sound in the
rhythm of the stimulus for visual and acoustic control.
If only the outer circle is visible, then the electrical stimulation circuit is not closed, i.e. there is no
stimulation current running through the patient.
If the circle is completely full, then the stimulation electrical circuit is closed, i.e. stimulation current
is running through the patient. The control sound will become more intensive in this case.
Due to a high stimulus voltage of max. 95 Vss , this renders an exceptionally large adjustment range for
extremely small stimulating electrodes.
Please make sure, that only CE-labeled adhesive electrodes are used for stimulation, which have been
provided with contact gel in order to keep the transition resistance as low as possible.
3.1 Technical data
Type:
MultiStim SWITCH
Type of device
BF
Battery:
9 V
Stimulating current:
max. 20 mAss/
Resistance range
0
⍀
– 60 k
⍀
Stimulation voltage:
max. 95 Vss
Stimulation frequency:
1 Hz/2 Hz
Stimulation impulse bandwidth:
0,05 ms / 0,10 ms / 0,20 ms/
0,30 ms/0,50 ms/1,00 ms
Operating conditions:
Temperature:
10°C – 30°C
Atmospheric humidity: 20% – 65%
Air pressure:
n.a.
Transportation- and storage conditions:
Temperature:
10°C – 30°C
Atmospheric humidity: 20% – 65%
Air pressure:
n.a.