15
Set-Up
NOTE
The microprocessor circuitry within the unit monitors the system to ensure proper
results. If any abnormalities are encountered, the Burn In cycle will be interrupted and
an error message displayed. If this situation should occur, turn the system off and
perform the procedure again. If the cycle is interrupted a second time, refer to Table
4 and check for handpiece malfunction. If a second handpiece is available, perform
the procedure using that handpiece.
Use the following instructions to perform the Heater Burn In procedure.
1. Place the Power Switch in the “OFF” ("0") position.
position.
5. Release the Program and Scroll Up (
) keys. The
Display will read “brn”.
6. Press the Scroll Up (
) Key to enter the Burn In
Mode. The handpiece heater will begin to heat
up with the temperature displayed (in °C or °F)
on the Display. The temperature of the heater
will stabilize at 315°C (or 600 °F) and remain at
that temperature for 10 minutes.
7. At the conclusion of the 10 minute period, the
heater temperature will increase to 427°C (or
800°F) and remain at that temperature for 15
minutes.
8. At the conclusion of the 15 minute time period, power is removed from the
heater. The Display will read “End”. Press and release the Scroll Up (
)
Key to exit Heater Burn In and return the to normal operation.
2. Ensure that the handpiece is connected to the system power source.
Some handpieces heater assemblies are shipped with a plastic cap
installed on the end of the heater assembly. If this cap is present, remove
the cap and discard. The cap is used for shipping purposes only.
3. Press and hold the Program and Scroll Up (
) keys together.
4. Place the Power Switch in the “ON” (
"I"
)
Содержание ST 115