Accessories
The accessories below are approved for use with OX-10
A
/5
A
Oxygen Concentrators:
Object/part No.
Type/model
Technical data
Bubble Humidifier(400 ml Capacity)
REF7100
ABS,PP
Cannula (disposable)
3583
PVC
Cannula w/ 2.1m(7 ft) Oxygen Tubing
(disposable)
3031
PVC
Oxygen Tubing (disposable)
30302
PVC
NOTE- The use of certain humidifiers and accessories not specified above may impair the oxygen
concentrator’s performance.
NOTE- If the “Service Required” light illuminates and the audible signal alarms but the unit is not
operating, there is no power to the unit. Refer to the Minor Troubleshooting chart on page 8 and
contact your OXTM Medical provider.
NOTE- from open to oxygen to reach the required performance required time should be within 5
minutes.
The OX-B series Device is a device within your concentrator that monitors the oxygen produced by your
unit. When the units is turned “ON,” all lights (Power/ Normal Oxygen, Low Oxygen, and Service
Required) on the front panel will illuminate briefly. After few seconds, only the “Power” and “Normal
Oxygen” lights will remain lit.
The lights on the front panel are defined as follows:
.
Green LED - Acceptable oxygen concentration
≥
82%.
.
Yellow LED - Below an acceptable oxygen concentration 82%~70%.
If the oxygen purity drop below the acceptable level, the Normal Oxygen light ( Green LED) will shut off
and the Low Oxygen light ( Yellow LED ) will illuminate. Switch to your reserve oxygen system. Refer to
the Troubleshooting section in this guide on page 8 and contact your OXTM provider.
As an added safety feature should the oxygen purity continue to drop, an intermittent audible signal will
alarm. Please contact your OXTM provider immediately. Do not attempt any other maintenance.
Page 11 of 16