8. When the pressure gauge shows 0 KPa, open the zippers and get out of the chamber

9.  After getting out of the chamber, turn off the air compressor

 

and Dehumidifier.

6. PREPARING BEFORE HYPERBARIC THERAPY

When using for the first time, Insure that the bubble wrap covering is removed from the Air deflate 
valve both internally and externally. Please test the empty chamber first and observe whether the 
equipment is operating normally. If the equipment is running abnormally, please do not use it.
Please contact your dealer promptly.

When you need to get out of the chamber, wait until the pressure gauge shows 0 KPa before you 
open the inner and outer zippers. Do not force the zipper to open during use or when pressure relief 
is not completed. 

In the event of an emergency or if the air deflate valve fails to operate properly, push the emergency 
relief valve from inside the chamber to relieve pressure. When the pressure is 0, pull the zipper and
get out of chamber. 

When you use the device for the first time, if there is slight discomfort or pain in your ears. This is a 
normal phenomenon caused by pressure increase. You can improve your discomfort by yawning,
chewing or swallowing. You can also slow pressurization by slightly opening the pressure relief valve 
to adapt to the pressure more gradually. After several uses, this phenomenon will gradually disappear. 

People may go to the chamber fully clothed, wear the loose, comfortable clothes.
Ladies should remove long socks(silk stocking) before entering into the chamber. Please take off shoes 
before getting in the chamber.
Shoes, jewelry, or any sharp objects that could damage the chamber should always be removed.
Disposable feet and hair covers may be used for increased sanitation.
To make the therapy experience more pleasant, the person should get as comfortable as possible, that 
means taking care of restroom breaks and/or drinking water before getting into the chamber.
The user should be advised to relax and breathe normally.
Breathe normally during the therapy. The air inside the chamber will be continually exchanged and people 
should have no difficulty breathing.

1. Clothing

or

External pressure gauge

Internal pressure gauge

11

ON/OFF

ON/OFF

Содержание Wheel Chair Ready Type Chamber

Страница 1: ...PO Box 2301 Naperville IL 60540 USA 630 452 8431 SalesTeam HyperbaricPro com www HyperbaricPro com AIR COMPRESSOR AIR DEHUMIDIFIER...

Страница 2: ...tem for the first time Keep this manual handy for your reference Please ensure that all items are removed from the box before trowing the box away Some customers have thrown away the protective paddin...

Страница 3: ...aintenance and Cultivation 8 FAQ 9 Faults and Troubleshooting 10 Condition for Transportation and Storage 4 4 5 6 7 9 11 12 13 14 14 CONTENTS Installation Working Scheme Accessories Air Compressor and...

Страница 4: ...e following table is a list of symbols and definitions that used with the Hyperbaric Chamber MILD HYPERBARIC OXYGEN THERAPY Authorized Representative In The European Community HOT Surface Do Not Touch...

Страница 5: ...people with severe infectious diseases and patients 2 Bad colds Pneumothorax patients 3 Patients with severe emphysema 4 Hypertension exceeds 160 100ml 5 Severe heart problems 6 Severe COPD Mild COPD...

Страница 6: ...de 10 Muffler Silencer 11 Blank Port reserved for future upgrades 12 Automatic Pressure Relief Valves 4 PRODUCT SPECIFICATIONS AND CONSTRUCTION 2 AIR COMPRESSOR Voltage 110 220V Air Flow 72 L Minute P...

Страница 7: ...LY AND OPERATION 4 Dehumidifier Voltage 110 220V Capability 150 L Minute Pressure 0 8mPa Power 300 Watt Weight 5kg 12 LPS Working Scheme Installation 1 Please open the accessory box before installatio...

Страница 8: ...Note Please install internal and external pressure gauges before use the internal pressure gauge should be installed inside the chamber external pressure gauge should be installed outside the chamber...

Страница 9: ...the oxygen generator 1 LPM below the maximum setting 2 Get into the chamber and pull outside zippers and then the inside zippers closed and make sure to flatten out the silicone strip between the zip...

Страница 10: ...ally and stablize the pressure betwen 30 KPa and 32 KPa 5 It is recommended to use 40 80 mins per session Pressure gauge error tolerance 3 KPa 40 80mins 6 After use please open the Air Deflate Valve t...

Страница 11: ...t time if there is slight discomfort or pain in your ears This is a normal phenomenon caused by pressure increase You can improve your discomfort by yawning chewing or swallowing You can also slow pre...

Страница 12: ...ure has stabilized When a session is finished unscrew the air deflate valve slowly to depressurize the chamber the pressure will drop slowly from 30 KPa to 0 KPa this depressurization process should b...

Страница 13: ...ssure and avoid the feeling of fullness in your ears you will need to clear your ears by chewing gum or food etc swallowing yawning During the pressurization of the chamber your mouth may become dry O...

Страница 14: ...h intake 3 Air compressor screws loose 4 Internal parts loose 1 Air tube is leaking at the joint 2 Didn t close the air deflate valve or didn t tighten up 3 Did not close one of the Zippers completely...

Страница 15: ...To help you choose the right Hyperbaric Chamber that fits your needs One of our hyperbaric specialists will assist you Please Call 630 451 8821...

Страница 16: ...This page intentionally left blank...

Отзывы: